Transcatheter Edge-to-Edge Repair in Patients With Anatomically Complex Degenerative Mitral Regurgitation.

Hausleiter Jörg, Lim D Scott, Gillam Linda D, Zahr Firas, Chadderdon Scott, Rassi Andrew N, Makkar Raj, Goldman Scott, Rudolph Volker, Hermiller James, Kipperman Robert M, Dhoble Abhijeet, Smalling Richard, Latib Azeem, Kodali Susheel K, Lazkani Mohamad, Choo Joseph, Lurz Philipp, O'Neill William W, Laham Roger, Rodés-Cabau Josep, Kar Saibal, Schofer Niklas, Whisenant Brian, Inglessis-Azuaje Ignacio, Baldus Stephan, Kapadia Samir, Koulogiannis Konstantinos, Marcoff Leo, Smith Robert L, PASCAL IID Registry Investigators

Многоцентровое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Mitral valve transcatheter edge-to-edge repair is safe and effective in treating degenerative mitral regurgitation (DMR) patients at prohibitive surgical risk, but outcomes in complex mitral valve anatomy patients vary.

OBJECTIVES

The PASCAL IID registry assessed safety, echocardiographic, and clinical outcomes with the PASCAL system in prohibitive risk patients with significant symptomatic DMR and complex mitral valve anatomy.

METHODS

Patients in the prospective, multicenter, single-arm registry had 3+ or 4+ DMR, were at prohibitive surgical risk, presented with complex anatomic features based on the MitraClip instructions for use, and were deemed suitable for the PASCAL system by a central screening committee. Enrolled patients were treated with the PASCAL system. Safety, effectiveness, and functional and quality-of-life outcomes were assessed. Study oversight also included an echocardiographic core laboratory and clinical events committee.

RESULTS

The study enrolled 98 patients (37.2% ≥2 independent significant jets, 15.0% severe bileaflet/multi scallop prolapse, 13.3% mitral valve orifice area <4.0 cm2, and 10.6% large flail gap and/or large flail width). The implant success rate was 92.9%. The 30-day composite major adverse event rate was 11.2%. At 6 months, 92.4% patients achieved MR ≤2+ and 56.1% achieved MR ≤1+ (P < 0.001 vs baseline). The Kaplan-Meier estimates for survival, freedom from major adverse events, and heart failure hospitalization at 6 months were 93.7%, 85.6%, and 92.6%, respectively. Patients experienced significant symptomatic improvement compared with baseline (P < 0.001).

CONCLUSIONS

The outcomes of the PASCAL IID registry establish the PASCAL system as a useful therapy for prohibitive surgical risk DMR patients with complex mitral valve anatomy. (PASCAL IID Registry within the Edwards PASCAL TrAnScatheter Valve RePair System Pivotal Clinical Trial [CLASP IID] NCT03706833).

Для цитирования: Hausleiter Jörg, Lim D Scott, Gillam Linda D, Zahr Firas, Chadderdon Scott, Rassi Andrew N, Makkar Raj, Goldman Scott, Rudolph Volker, Hermiller James, Kipperman Robert M, Dhoble Abhijeet, Smalling Richard, Latib Azeem, Kodali Susheel K, Lazkani Mohamad, Choo Joseph, Lurz Philipp, O'Neill William W, Laham Roger, Rodés-Cabau Josep, Kar Saibal, Schofer Niklas, Whisenant Brian, Inglessis-Azuaje Ignacio, Baldus Stephan, Kapadia Samir, Koulogiannis Konstantinos, Marcoff Leo, Smith Robert L, PASCAL IID Registry Investigators. Transcatheter Edge-to-Edge Repair in Patients With Anatomically Complex Degenerative Mitral Regurgitation.. Journal of the American College of Cardiology. 2023. DOI: 10.1016/j.jacc.2022.11.034.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.