Mauri Victor, Sugiura Atsushi, Spieker Max, Iliadis Christos, Horn Patrick, Öztürk Can, Besler Christian, Riebisch Matthias, Al-Hammadi Osamah, Ruf Tobias, Gerçek Muhammed, Grothusen Christina, Mehr Michael, Becher Marc Ulrich, Mues Christoph, Boeder Niklas, Kreidel Felix, Friedrichs Kai, Westenfeld Ralf, Braun Daniel, Baldus Stephan, Rassaf Tienush, Thiele Holger, Nickenig Georg, Hausleiter Jörg, Möllmann Helge, Kelm Malte, Rudolph Volker, von Bardeleben Ralph Stephan, Nef Holger M, Luedike Peter, Lurz Philipp, Pfister Roman

Многоцентровое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

In addition to the edge-to-edge MitraClip repair system, the edge-to-spacer PASCAL repair system was approved for percutaneous treatment of severe mitral regurgitation (MR). Comparative data are lacking.

OBJECTIVES

The aim of this study was to compare procedural and short-term safety and efficacy of 2 leaflet-based transcatheter mitral valve repair systems.

METHODS

Procedural and 30-day outcomes were investigated in a propensity score-matched cohort of 307 PASCAL and 307 MitraClip patients at 10 sites. Matching criteria included sex, age, left ventricular ejection fraction, New York Heart Association functional class, MR etiology, left ventricular end-diastolic diameter, left atrial volume index, and vena contracta width. The primary efficacy endpoints were technical success and degree of residual MR at discharge. The primary safety endpoint was the rate of major adverse events (MAE).

RESULTS

Technical success was 97.0% in the PASCAL group and 98.0% in the MitraClip group (P = 0.624). MR ≤2+ at discharge was comparable in both groups (PASCAL: 93.8% vs MitraClip: 92.4%; P = 0.527), with more patients exhibiting MR ≤1+ in the PASCAL group (70.5% vs 56.6%; P < 0.001). The postprocedural mean gradient was significantly higher in the MitraClip group (3.3 ± 1.5 mm Hg vs 3.9 ± 1.7 mm Hg; P < 0.001). At 30 days, all-cause mortality and MAE rates were similar (mortality: 1.7% vs 3.3%; P = 0.299; MAE: 3.9% vs 5.2%; P = 0.562).

CONCLUSIONS

In this first large propensity score-matched comparison, procedural success rates and MAE did not differ significantly between patients treated with the PASCAL or MitraClip valve repair system. Procedural results with less than moderate MR and no elevated transmitral gradient were more common in the PASCAL group, which might have an impact on long-term outcome.

Для цитирования: Mauri Victor, Sugiura Atsushi, Spieker Max, Iliadis Christos, Horn Patrick, Öztürk Can, Besler Christian, Riebisch Matthias, Al-Hammadi Osamah, Ruf Tobias, Gerçek Muhammed, Grothusen Christina, Mehr Michael, Becher Marc Ulrich, Mues Christoph, Boeder Niklas, Kreidel Felix, Friedrichs Kai, Westenfeld Ralf, Braun Daniel, Baldus Stephan, Rassaf Tienush, Thiele Holger, Nickenig Georg, Hausleiter Jörg, Möllmann Helge, Kelm Malte, Rudolph Volker, von Bardeleben Ralph Stephan, Nef Holger M, Luedike Peter, Lurz Philipp, Pfister Roman. Early Outcomes of 2 Mitral Valve Transcatheter Leaflet Approximation Devices: A Propensity Score-Matched Multicenter Comparison.. JACC. Cardiovascular interventions. 2022. DOI: 10.1016/j.jcin.2022.10.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.