MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.30490

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2022

A novel comprehensive radiation shielding system eliminates need for personal lead aprons in the catheterization laboratory.

Rabah Maher, Allen Sorcha, Abbas Amr E, Dixon Simon

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

This clinical study evaluated the efficacy of a novel radiation shielding system for the cardiac catheterization laboratory designed to provide comprehensive protection that obviates the need for personal lead aprons.

BACKGROUND

Invasive Cardiologists are exposed to occupational health hazards related directly to radiation exposure (RE) and indirectly to the orthopedic burden of wearing only partially protective lead aprons. Innovations to reduce these risks are warranted. A novel comprehensive shielding system (ProtegoTM , Image Diagnostics Inc, Fitchburg, Ma) has been validated in pre-clinical studies to provide excellent radiation protection, sufficient for the State of Michigan to certify it for use without need for personal lead aprons.

METHODS

This clinical analysis measured RE to a single Physician operator utilizing the ProtegoTM shield (and not wearing personal lead apron) during routine cardiac catheterization procedures (diagnostic and interventional). RE was measured at both thyroid and waist level with a real-time dosimetry system (RaysafeTM , Billdal, Sweden), calculated on a median per case basis (mrems). Additional parameters collected included procedure type, access site, per case fluoroscopy time, and patient factors including body mass index.

RESULTS

In n=98 cases (25% diagnostic, 75% interventional including 22% chronic total occlusions), median/case RE was 0.4 mrems (thyroid) and 0.2 mrems (waist). RE=0 in 12 cases. In no case did radiation exposure exceed 3.2 mrems.

CONCLUSION

The ProtegoTM shield system provides excellent RE protection to the Physician operator, without the need for personal lead aprons and has the potential to reduce catheterization laboratory occupational health hazards.

Для цитирования: Rabah Maher, Allen Sorcha, Abbas Amr E, Dixon Simon. A novel comprehensive radiation shielding system eliminates need for personal lead aprons in the catheterization laboratory.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2022. DOI: 10.1002/ccd.30490.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.