DOI 10.1016/j.jcin.2022.08.034
Evaluation of Multimodality LAA Leak Closure Methods Following Incomplete Occlusion: The LAA Leak Study.
Аннотация
BACKGROUND
Incomplete left atrial appendage (LAA) closure is an evolving topic of clinical significance and thromboembolic potential, with recent long-term studies suggesting lower cutoffs for relevant leak size.
OBJECTIVES
The aim of this prospective observational study was to assess 3 different closure techniques for persistent peridevice leaks after incomplete LAA closure and compare their efficacy and safety outcomes.
METHODS
We studied 160 patients (mean age 72 ± 9 years; 71% men) who underwent 1 of the 3 available modalities (detachable embolization coils, vascular plugs or septal occluders, and radiofrequency ablation) for residual central or eccentric leak closure. Both acute postprocedural success (closure or <1-mm leak at the end of the procedure) and closure at 1-year follow-up transesophageal echocardiography imaging were evaluated.
RESULTS
Of 160 patients, 0.6%, 41.3%, and 58.1% had mild (1-2 mm), moderate (3-5 mm), and severe (≥5 mm) leaks, respectively. Baseline LAA closure type was 72.5% Watchman FLX, 16.3% Lariat, 5.6% surgical ligation, 1.9% AtriClip, and 1.9% Amulet. Successful closure (0- or <1-mm leak) was seen in 100% of patients in all cohorts following intervention, with overall complete closure (0-1 mm) or mild or minimal leaks (1-2 mm) on 1-year follow-up transesophageal echocardiography seen in 100% of the atrial septal occluder or vascular plug cohort, 85.9% of the coil cohort, and 83.3% of the radiofrequency ablation cohort (P < 0.001). Two patients (1.3%) experienced cardiac tamponade, and there were no deaths or other complications.
CONCLUSIONS
Peridevice leaks can safely and effectively be closed using 3 different modalities depending on size and location.