MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00330-022-08938-6

Медицинская библиотека · European radiology · 2022

First-pass perfusion cardiovascular magnetic resonance parameters as surrogate markers for left ventricular diastolic dysfunction: a validation against cardiac catheterization.

Guo Xinli, Gong Chao, Song Rizhen, Wan Ke, Han Yuchi, Chen Yucheng

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

The non-invasive assessment of left ventricular (LV) diastolic dysfunction remains a challenge. The role of first-pass perfusion cardiac magnetic resonance (CMR) parameters in quantitative hemodynamic analyses has been reported. We therefore aimed to validate the diagnostic ability and accuracy of such parameters against cardiac catheterization for LV diastolic dysfunction in patients with left heart disease (LHD).

METHODS

We retrospectively enrolled 77 LHD patients who underwent CMR imaging and cardiac catheterization. LV diastolic dysfunction was defined as pulmonary capillary wedge pressure (PCWP) or LV end-diastolic pressure (LVEDP) > 12 mmHg on catheterization. On first-pass perfusion CMR imaging, pulmonary transit time (PTT) was measured as the time for blood to pass from the left ventricle to the right ventricle (RV) through the pulmonary vasculature. Pulmonary transit beat (PTB) was the number of cardiac cycles within the interval, and pulmonary blood volume indexed to body surface area (PBVi) was the product of PTB and RV stroke volume index (RVSVi).

RESULTS

Of the 77 LHD patients, 53 (68.83%) were found to have LV diastolic dysfunction, and showed significantly higher PTTc, PTB, and PBVi (p < 0.05) compared with those without. In multivariate analyses, only PTTc and PTB were identified as independent predictors of LV diastolic dysfunction (p < 0.05). With an optimal cutoff of 11.9 s, PTTc yielded the best diagnostic performance for LV diastolic dysfunction (area under the curve = 0.83, p < 0.001).

CONCLUSIONS

PTTc may represent a non-invasive quantitative surrogate marker for the detection and assessment of diastolic dysfunction in LHD patients.

KEY POINTS: • PTTc yielded the best diagnostic accuracy for diastolic dysfunction, with an optimal cutoff of 11.9 s, and a specificity of 100%. • PTTc and PTB were found to be independent predictors of LV diastolic dysfunction across different multivariate models with high reproducibility. • PTTc is a promising non-invasive surrogate marker for the detection and assessment of diastolic dysfunction in LHD patients.

Для цитирования: Guo Xinli, Gong Chao, Song Rizhen, Wan Ke, Han Yuchi, Chen Yucheng. First-pass perfusion cardiovascular magnetic resonance parameters as surrogate markers for left ventricular diastolic dysfunction: a validation against cardiac catheterization.. European radiology. 2022. DOI: 10.1007/s00330-022-08938-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.