MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.carrev.2022.05.009

Медицинская библиотека · Cardiovascular revascularization medicine : including molecular interventions · 2022

Case Report: WATCHMAN Procedure Complicated By Atrial Perforation and Device Dislodgement.

Al Rameni Dina, de Armas Ismael Salas, Al Rameeni Aon, Gupta Sameer, Gregoric Igor D

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Atrial perforation and device dislodgment are uncommon encounters after percutaneous left atrial appendage exclusion procedures. The literature describes how various endovascular and surgical approaches can be applied to treat these complications. In this case report, we describe an atrial perforation that occurred after a WATCHMAN™ FLX device (Boston Scientific) deployment. Heart manipulation with lifting was required for the atrial repair. During this maneuver, the device dislodged and was managed by surgical device extraction and left atrial appendage resection and suture closure. Minimal to no manipulation of the left atrium during surgical exploration was found to be invaluable in order to avoid device dislodgement and additional surgical corrective interventions. Additionally, intraoperative transesophageal echocardiography at the beginning and conclusion of the case is key in monitoring the device position and identifying device dislodgement.

Для цитирования: Al Rameni Dina, de Armas Ismael Salas, Al Rameeni Aon, Gupta Sameer, Gregoric Igor D. Case Report: WATCHMAN Procedure Complicated By Atrial Perforation and Device Dislodgement.. Cardiovascular revascularization medicine : including molecular interventions. 2022. DOI: 10.1016/j.carrev.2022.05.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.