MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.30237

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2022

Real-world experience with the new Watchman FLX device: Data from two high-volume Sicilian centers. The FLX-iEST registry.

Vizzari Giampiero, Grasso Carmelo, Sardone Andrea, Mazzone Paolo, Laterra Giulia, Frazzetto Marco, Sacchetta Giorgio, Micari Antonio, Tamburino Corrado, Contarini Marco

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The Watchman-FLX left atrial appendage closure (LAAC) device presents innovative features: higher conformability, reduced length, closed distal "flex-ball" during deployment, and flattened surface. We report our real-world experience with the Watchman-FLX device in two centers with consolidated LAAC expertise.

METHODS

We enrolled 200 consecutive Watchman-FLX patients (2019-2021) in a nonrandomized double-center registry; procedural data and follow-up for midterm clinical outcomes were collected. A control group of 100 patients treated with first-generation Watchman (2.5) was included.

RESULTS

According to mean CHAD2 DS2 -VASc (5 ± 1.40) and HAS-BLED (3.8 ± 1.01) scores, the population included in this study was at high risk: 29% had a previous stroke and 56.5% a bleeding event. Main LAAC indications were symptomatic hemorrhage (39.5%), need for triple antithrombotic therapy (39%), gastrointestinal bleeding (32%), and oral anticoagulation intolerance (18%). Transesophageal echocardiography guidance was followed in 93% of cases (48% in general anesthesia and 45% under conscious sedation). Repositioning an FLX device was required in 20% of cases and no complication occurred. In 96% of patients, the first selected device was delivered, while in 4% a device size change was required after the first choice (7% with Watchman 2.5). Peridevice leaks (<5 mm) were found postimplant in two cases (1%). Overall, the procedural success rate was 99.5%. One patient's procedure was unsuccessful (0.5%), due to left atrial appendage (LAA) anatomy; differently, the mean failure rate with Watchman 2.5 was 2%. No device embolization was reported. Complications (8.5%) were mainly related to the access site (3%); major bleedings (1%), and in-hospital death (0.5%) rarely occurred. After a follow-up of 272 ± 173 days, 2.3% of cases experienced a non-device-related stroke and 0.6% fatal bleeding.

CONCLUSION

Our registry showed a high procedural success rate of the Watchman-FLX in a high-risk population. According to our experience, the main advantages include easy implanting and repositioning, absence of embolization, good LAA sealing, and low rate of complications in the follow-up period.

Для цитирования: Vizzari Giampiero, Grasso Carmelo, Sardone Andrea, Mazzone Paolo, Laterra Giulia, Frazzetto Marco, Sacchetta Giorgio, Micari Antonio, Tamburino Corrado, Contarini Marco. Real-world experience with the new Watchman FLX device: Data from two high-volume Sicilian centers. The FLX-iEST registry.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2022. DOI: 10.1002/ccd.30237.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.