DOI 10.1002/ccd.30178

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2022

Temporal changes and clinical significance of peridevice leak following left atrial appendage occlusion with Amplatzer devices.

Korsholm Kasper, Jensen Jesper M, Nørgaard Bjarne L, Nielsen-Kudsk Jens E

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The natural history of peridevice leak (PDL) following left atrial appendage occlusion (LAAO) is unknown. This study sought to investigate changes of PDL from 2 until 12 months after LAAO, using cardiac computed tomography (CT), and to assess the potential association between persistent PDL and clinical outcomes

METHODS

Single-center observational study of Amplatzer LAAO implants between 2010 and 2017 (n = 206). Patients with 2 and 12 months cardiac CT were included in the study (n = 153). Images were blindly analyzed. PDL was characterized by frequency and size at the device disc, lobe, and left atrial appendage contrast patency. Patients were followed for the composite outcome of ischemic stroke, transient ischemic attack, systemic embolism, or all-cause death. Median follow up from LAAO was 3.1 (2.3-4.3) years.

RESULTS

Contrast patency was present in 101 (66%) and 72 (47%) (p < 0.001) at 2 and 12 months, respectively. PDL was identified at the disc in 103 (67%) patients at 2 months versus 93 (61%) at 12 months (p = 0.08), and at the lobe in 29 (19%) at both time points. PDL area at the disc did not change significantly over time, ∆$\unicode{x02206}$ area: -8.95 mm (95% confidence interval [CI]: -18.9; 1.01) p = 0.08. Permanent atrial fibrillation was independently associated with persistent PDL. Persistent versus no PDL was associated with a 62% worse clinical outcome, however not statistically significant, hazard ratio (HR): 1.62 (95% CI: 0.9-2.93), p = 0.11.

CONCLUSION

Persistent PDL was frequently observed following LAAO with Amplatzer devices. The PDL frequency and size appeared unchanged between 2 and 12 months. Persistent PDL was not significantly associated with worse clinical outcomes, yet this needs further delineation in future studies.

Для цитирования: Korsholm Kasper, Jensen Jesper M, Nørgaard Bjarne L, Nielsen-Kudsk Jens E. Temporal changes and clinical significance of peridevice leak following left atrial appendage occlusion with Amplatzer devices.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2022. DOI: 10.1002/ccd.30178.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.