Objective

The purpose of this case study was to demonstrate hearing preservation of a subject who was implanted with a 10-mm short electrode cochlear implant that was determined to be malfunctioning at 6 months postimplantation and was explanted and reimplanted with a 16-mm short electrode device.Study

Design

Single-subject case study.

Setting

Research hospital.Patient: A 60-year old female with a history of gradual progressive bilateral steeply sloping sensorineural hearing loss.Intervention: Rehabilitative.Main Outcome Measure(s): Audiometric data and speech perception in quiet and in noise were collected pre- and postoperatively at 3 and 6 months with the 10-mm short electrode device and pre-explantation and postoperatively at 3, 6, and 12 months with the 16-mm short electrode device.

Results

Functional hearing preservation was accomplished following surgical implantation of both short electrode devices. Overall, the subject had a 22 dB HL total shift in pure-tone-average (0.125–1 kHz) after two cochlear implant surgeries. Speech perception growth was limited over the 6 months the subject was implanted with the 10-mm short electrode device. After 3 months of experience with the 16-mm short electrode device, the subject experienced significant improvements in both speech perception in quiet and in noise.

Conclusions

The inner ear might be more robust than once thought, as was determined through preservation of residual hearing after implantation of two hearing preservation cochlear implants and one device explantation. Furthermore, it is important that hearing professionals remain cognizant of unusual speech perception patterns associated with the cochlear implant. Key Words: Hearing preservation—Residual hearing—Short electode. [Corrected]

Для цитирования: Dunn Camille C, Etler Christine, Hansen Marlan, Gantz Bruce J. Successful Hearing Preservation After Reimplantation of a Failed Hybrid Cochlear Implant.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2015. DOI: 10.1097/mao.0000000000000867.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.