Медицинская библиотека · Journal of sport rehabilitation · 2013

Test-retest reliability of biodex system 4 pro for isometric ankle-eversion and -inversion measurement.

Tankevicius Gediminas, Lankaite Doanata, Krisciunas Aleksandras

СтатьяПеревод пока не подготовлен

Аннотация

CONTEXT: The lack of knowledge about isometric ankle testing indicates the need for research in this area.

OBJECTIVE

to assess test-retest reliability and to determine the optimal position for isometric ankle-eversion and -inversion testing.

DESIGN

Test-retest reliability study. Isometric ankle eversion and inversion were assessed in 3 different dynamometer foot-plate positions: 0°, 7°, and 14° of inversion. Two maximal repetitions were performed at each angle. Both limbs were tested (40 ankles in total). The test was performed 2 times with a period of 7 d between the tests.

SETTING

University hospital.

PARTICIPANTS

The study was carried out on 20 healthy athletes with no history of ankle sprains.

MAIN OUTCOME MEASURES: Reliability was assessed using intraclass correlation coefficient (ICC2,1); minimal detectable change (MDC) was calculated using a 95% confidence interval. Paired t test was used to measure statistically significant changes, and P <.05 was considered statistically significant.

RESULTS

Eversion and inversion peak torques showed high ICCs in all 3 angles (ICC values .87-.96, MDC values 3.09-6.81 Nm). Eversion peak torque was the smallest when testing at the 0° angle and gradually increased, reaching maximum values at 14° angle. The increase of eversion peak torque was statistically significant at 7 ° and 14° of inversion. Inversion peak torque showed an opposite pattern-it was the smallest when measured at the 14° angle and increased at the other 2 angles; statistically significant changes were seen only between measures taken at 0° and 14°.

CONCLUSIONS

Isometric eversion and inversion testing using the Biodex 4 Pro system is a reliable method. The authors suggest that the angle of 7° of inversion is the best for isometric eversion and inversion testing.

Для цитирования: Tankevicius Gediminas, Lankaite Doanata, Krisciunas Aleksandras. Test-retest reliability of biodex system 4 pro for isometric ankle-eversion and -inversion measurement.. Journal of sport rehabilitation. 2013. DOI: 10.1123/jsr.22.3.212.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.