The presacral space and its impact on sacral neuromodulator implantation.
Аннотация
PURPOSE: We describe the presacral space and its potential impact on sacral neuromodulator implantation and bowel injury.
MATERIALS AND
METHODS
Parasagittal images containing bilateral sacral foramina (S2-S4) were examined on 45 pelvic magnetic resonance images. Images were excluded from analysis if they were poor quality or had any history causing distortion of normal anatomy. We measured the natural angle between the foramina and the dorsal skin to approximate the needle angulation during neuromodulator electrode placement. Using these angles we measured the distance from the skin to any bowel (D1), the skin to the dorsal sacrum (D2) and then calculated the distance from the dorsal sacrum to any bowel (D3).
RESULTS
Mean subject age was 45 years (range 19 to 78) and body mass index was 27.9 kg/m(2) (range 18.6 to 56.2). At S3 the mean foraminal angle and D3 were 46 +/- 8.4 degrees and 27.4 +/- 11.7 mm, respectively. Increasing age was moderately correlated to widening D3 at each foramina (r = 0.3, Pearson's p <0.05). Body mass index did not consistently vary with D3 at any foramina.
CONCLUSIONS
Our measurements suggest that the presacral space can be expected to be approximately 27 mm at the level of S3 where the neuromodulator electrode is implanted. It is possible to encounter bowel while performing this implantation using standard techniques and equipment. We recommend the standard use of fluoroscopy during placement.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.