OBJECTIVE

The purpose of our study was to develop a method to subtract barium-labeled stool from the colon using a phantom and to evaluate the performance of the technique in a pilot human population.

MATERIALS AND

METHODS

A phantom containing 6-mm flat polyps and three types of simulated stool (homogeneous, moderately heterogeneous, and severely heterogeneous) mixed with barium was created, scanned, and tested using three stool subtraction algorithms but no cathartic. Thirty patients with suspected colorectal polyps were studied using stool tagging to determine which was the most effective stool subtraction algorithm. Colonoscopy was the reference standard. Examinations were evaluated blindly using the unsubtracted and 6 weeks later both the unsubtracted and subtracted data sets.

RESULTS

A threshold of 200 H and expansion and convolution techniques were the most effective tools for subtracting stool and minimizing artifacts. When applied to the human population, sensitivities using the unsubtracted data sets were 90% (18/20) and 68% (26/38) for polyps > or = 1 cm and > or = 5 mm, respectively. Specificities were 100% (4/4) and 75% (3/4) for polyps > or = 1 cm and > or = 5 mm. For the stool-subtracted data sets, sensitivities were 90% (18/20) and 71% (27/38) for polyps > or = 1 cm and > or = 5 mm. Per patient sensitivities were 88% (15/17) and 77% (20/26) for > or = 1 cm and > or = 5 mm polyps. Specificities were 100% (4/4) for large polyps and 25% (1/4) for smaller polyps.

CONCLUSION

Image processing tools combining thresholding, expansion, and convolution were the most useful for stool subtraction. Laxative-free colon examinations using barium for stool labeling can be performed at CT colonography with or without stool subtraction with high accuracy. Further study is warranted.

Для цитирования: Johnson Kristina T, Carston Michael J, Wentz Robert J, Manduca Armando, Anderson Steven M, Johnson C Daniel. Development of a cathartic-free colorectal cancer screening test using virtual colonoscopy: a feasibility study.. AJR. American journal of roentgenology. 2007. DOI: 10.2214/ajr.05.1484.