Dose reduction in gastrointestinal and genitourinary fluoroscopy: use of grid-controlled pulsed fluoroscopy.

Boland G W, Murphy B, Arellano R, Niklason L, Mueller P R

Клиническое исследованиеСравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

We evaluated the diagnostic accuracy of a grid-controlled fluoroscopy unit compared with a conventional continuous fluoroscopy unit for a variety of abdominal and pelvic fluoroscopic examinations.

SUBJECTS AND

METHODS

Seventy patients (29 men and 41 women; age range, 24-78 years) were enrolled in one of seven abdominal and pelvic fluoroscopic examinations, including upper gastrointestinal series (n = 20), barium enema (n = 10), voiding cystourethrogram (n = 10), percutaneous abdominal catheter tube injection (n = 10), hysterosalpingogram (n = 10), and percutaneous needle insertion and catheter placement (nephrostomy, percutaneous biliary drainage) (n = 10). Each patient underwent at least 10 sec of continuous fluoroscopy that was randomly and blindly compared with 10-sec periods of pulsed fluoroscopy at 15, 7.5, and 3.75 frames per second. A radiologist outside the examination room, unaware of the frame rate per second, evaluated the procedure in real time on a television monitor. The radiologist assessed image quality and diagnostic acceptability using a scoring system. Statistical analysis was performed using the paired Student's t test.

RESULTS

For all procedures at all frame rates, we found no statistically significant superiority of one frame rate over another. For most procedures, the slower frame rates were considered equivalent to continuous fluoroscopy when the images were assessed for image quality and diagnostic confidence.

CONCLUSION

Our findings suggest that most abdominal and pelvic fluoroscopic procedures can be performed at substantially lower frame rates than those used for continuous fluoroscopy; adopting this procedure may lead to substantial dose savings for the patient and the fluoroscopy operator.

Для цитирования: Boland G W, Murphy B, Arellano R, Niklason L, Mueller P R. Dose reduction in gastrointestinal and genitourinary fluoroscopy: use of grid-controlled pulsed fluoroscopy.. AJR. American journal of roentgenology. 2000. DOI: 10.2214/ajr.175.5.1751453.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.