DOI 10.1002/jmri.70575

Медицинская библиотека · Journal of magnetic resonance imaging : JMRI · 2026

Multicenter, Multivendor Development and Validation of Automated Liver Prescription.

Fullerton Garrett C, Starekova Jitka, Buelo Collin J, Harris David T, do Vale Souza Raphael, Faacks Aaron, Anagnostopoulos Alexandra A, Christofi Veronica P, Murphy Austin, Kadi Diana, Yokoo Takeshi, Bashir Mustafa R, Reeder Scott B, Hernando Diego

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Manual prescription of liver MRI volumes introduces inter-operator variability and prolongs exam duration. Artificial intelligence (AI)-based prescription has been demonstrated, but existing work has been limited to single-vendor settings.

PURPOSE: To develop and validate an AI model for automated liver MRI prescription using a large, consecutive, multicenter dataset encompassing multiple vendors, field strengths, localizer sequences, and liver pathologies.

STUDY TYPE: Retrospective.

POPULATION: A total of 11,012 patients (5928 female) who underwent liver MRI exams across three vendors (GE HealthCare, Philips Healthcare, Siemens Healthineers) at three institutions, split into training/validation/test sets (70/10/20%). Liver disorders included cirrhosis, ascites, portal hypertension, and iron overload.

FIELD STRENGTH/SEQUENCE: Three-plane localizers acquired using single-shot fast spin echo (68.3%) and gradient echo (31.7%) at 1.5 T (52.8%) and 3 T (47.2%).

ASSESSMENT: A YOLOv8 object detection model was trained to detect the liver, torso, and arms on localizer images. Automated prescription accuracy was assessed using three-dimensional intersection-over-union (IoU3D) for axial, coronal, and sagittal prescriptions. Subgroup analyses evaluated performance across technical and patient-related factors.

STATISTICAL TESTS: Mann-Whitney U tests with Bonferroni correction, Kruskal-Wallis tests, Wilcoxon signed-rank tests, Spearman correlation. Effect sizes were calculated using rank-biserial correlation (rrb) and epsilon-squared (ε2). p < 0.05 was considered statistically significant.

RESULTS

Median axial prescription IoU3D was 0.963 (interquartile range: 0.946-0.976), with median IoU3D above 0.940 for coronal and sagittal prescriptions. For axial prescriptions, margins less than 4 mm in each direction were required to achieve complete coverage of manual prescriptions in 90% of cases in the held-out test set. Effects were negligible to small across vendors, field strengths, localizer sequences, and patient factors (rrb ≤ 0.16, ε2 ≤ 0.01).

DATA

CONCLUSION

A single multivendor model achieved accurate automated liver MRI prescription without clinically meaningful performance differences across vendors, field strengths, localizer sequences, or patient factors.

EVIDENCE LEVEL: 2.

TECHNICAL EFFICACY: Stage 1.

Для цитирования: Fullerton Garrett C, Starekova Jitka, Buelo Collin J, Harris David T, do Vale Souza Raphael, Faacks Aaron, Anagnostopoulos Alexandra A, Christofi Veronica P, Murphy Austin, Kadi Diana, Yokoo Takeshi, Bashir Mustafa R, Reeder Scott B, Hernando Diego. Multicenter, Multivendor Development and Validation of Automated Liver Prescription.. Journal of magnetic resonance imaging : JMRI. 2026. DOI: 10.1002/jmri.70575.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.