DOI 10.1016/j.jaccas.2025.104559
Clip Liberation to Facilitate Transcatheter Tricuspid Valve Replacement (CLEFT).
Аннотация
BACKGROUND
Transcatheter tricuspid valve edge-to-edge repair (T-TEER) can be effective for patients with tricuspid regurgitation (TR), but TR may recur, requiring transcatheter tricuspid valve replacement (TTVR). Although TTVR has been described in patients with single leaflet device attachment of edge-to-edge devices, implantation in presence of well-attached devices has not been described.
CASE SUMMARY: We present a patient with recurrent TR after T-TEER. Transesophageal echocardiography revealed that 3 previously placed MitraClip XTW devices were well attached to the tricuspid leaflets, with recurrent severe TR. After laceration of 1 leaflet attached to the central XTW clip with transcatheter electrosurgery, we implanted an EVOQUE TTVR with trivial paravalvular leakage.
DISCUSSION: Transcatheter electrosurgery can free a tricuspid clip from leaflets, allowing implantation of an EVOQUE TTVR with excellent TR resolution.
TAKE-HOME MESSAGE: Transcatheter electrosurgical liberation of T-TEER devices might be a viable technique in patients who require a TTVR for recurrent tricuspid regurgitation.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.