A comparative study in patients with type 2 von Willebrand disease using 4 different platelet-dependent von Willebrand factor assays.
Аннотация
BACKGROUND
Several assays are now available to evaluate platelet-dependent von Willebrand factor (VWF) activity.
OBJECTIVE
To report the results obtained using 4 different assays in patients with von Willebrand disease (VWD) carrying variants mainly in the A1 domain, which is critical for VWF binding to glycoprotein Ib (GPIb) and ristocetin.
METHODS
We evaluated 4 different assays, 2 gain-of-function mutant GPIb binding (VWF:GPIbM) and 2 ristocetin cofactor (VWF:RCo) assays, in 76 patients with type 2 VWD. Patients and healthy controls were tested using VWF:GPIbM enzyme-linked immunosorbent assay (ELISA), VWF:GPIbM automated, VWF:RCo aggregometric, and VWF:RCo automated assays.
RESULTS
There was a good correlation (Pearson's r>0.82) and agreement (Bland-Altman plots assessment) between the 4 assays, although several outliers existed among the type 2B without high-molecular-weight multimers (HMWM). The VWF activity/VWF:antigen ratios, calculated for each assay, were used to establish the percentage of a correct diagnosis of type 2 (ratio<0.60) in these patients: VWF:RCo aggregometric, 2A(100%), 2M(78%), 2M/2A(100%), 2B(68%); VWF:RCo automated, 2A(88%), 2M(89%), 2M/2A(100%), 2B(63%); VWF:GPIbM ELISA, 2A(96%), 2M(67%), 2M/2A(67%), 2B(0%); VWF:GPIbM automated, 2A(73%), 2M(44%), 2M/2A(75%), 2B(84%). In type 2B patients with HMWM, all assays gave a ratio ≥0.60.
CONCLUSION
The VWF:GPIbM-automated assay is the most effective to diagnose as type 2 the 2B variants, whereas the VWF:RCo assays are the most effective in detecting 2M and 2M/2A variants. The VWF:GPIbM ELISA greatly overestimates the activity of the type 2B patients lacking HMWM. In this study, the use of a VWF activity/VWF:antigen ratio cut-off of 0.70 halved the number of misdiagnosed patients.