Cortical bone trajectory fixation cause low compression force in anterior vertebral column.

Fujiwara Sho, Ohnishi Yuichiro, Iwatsuki Koichi, Kishima Haruhiko

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The cortical bone trajectory (CBT) screws follow a caudocephalad and lateral path from the pedicle to the vertebral body. The bone fusion rate of CBT fixation is equal to or lower than conventional pedicle screw fixation. It remains unclear whether or not CVT screws exert equally compressive forces across the vertebral column. In this study, we intraoperatively examined the insertional torque of CBT screws, and investigated the compression loading and pressure distribution in conventional and CBT fixation using pig bones.

METHODS

The insertional torque was measured for a total of 115 CBT screws. Detailed positions of these screws were retrospectively confirmed using CT scans. Screw loosening and interbody fusion were examined 1 year after surgery. In the experiment using pig bones, we inserted screws by conventional trajectory (n = 3) and CBT (n = 4).

RESULTS

Multiple regression analysis showed that the total screw length, the distance from the screw to the medial border of the pedicle and the distance from the superior endplate of the vertebrae were significant independent factors affecting the insertional torque. There was no significant association between the insertional torque and the radiographic bone fusion rate 12 months after surgery. The average pressure and the compression loading in the anterior column were significantly lower in CBT group.

CONCLUSION

These results suggested that the long CBT screws as close to the endplate had high insertional torque, but the anterior column in CBT fixation showed low compression force leading to the insufficient holding intervertebral cage. CBT screws may cause the micromotion of cages, which lowers the bone fusion rate.

Для цитирования: Fujiwara Sho, Ohnishi Yuichiro, Iwatsuki Koichi, Kishima Haruhiko. Cortical bone trajectory fixation cause low compression force in anterior vertebral column.. North American Spine Society journal. 2022. DOI: 10.1016/j.xnsj.2022.100113.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.