Bespoke valve sizing avoids annular rupture in patients treated with a balloon-expandable transcatheter heart valve.

Chavarria Jorge, Falcao Felipe, Dutra Gustavo, Sathananthan Janarthanan, Wood David, Douvi Georgia, Natarajan Madhu, Jaffer Iqbal, Velianou James, Sheth Tej

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Annular and left ventricular outflow tract (LVOT) calcification increase the risk of annular rupture following transcatheter aortic valve replacement (TAVR). The outcomes of a strategy of routine use of a balloon-expandable valve (BEV) for all patients irrespective of annular or LVOT calcium is unknown.

OBJECTIVES

We evaluated the impact of bespoke sizing on annular rupture in patients treated with a BEV.

METHODS

All consecutive patients undergoing TAVR at a single centre (February 2020-February 2022) were treated only with a BEV. No other valve design was used. Annular/LVOT calcification was assessed using a standardized grading system. For each annular area, we determined the percentage valve oversizing with nominal deployment. The balloon deployment volume was then adjusted when required (over-/underfilled) to achieve over-sizing of approximately 5% in the presence of annular/LVOT calcium and 5%-10% in the absence of annular/LVOT calcium. Adjusted valve areas were assumed to change proportionately to the change in balloon deployment volume.

RESULTS

Among 533 TAVR treated patients, annular/LVOT calcification was present in 166 (31.1%) and moderate or severe in 90 (16.9%). In patients with annular/LVOT calcification, the adjusted oversizing was 3.5 ± 3.6% and in patients without annular/LVOT calcification, the adjusted oversizing was 6.8 ± 4.7% (p < 0.001). There were no cases of annular rupture and no cases with more than mild paravalvular leak (PVL). Mild PVL was more frequent in patients with annular/LVOT calcium (10.8% vs 4.6%, p = 0.01).

CONCLUSION

Bespoke BEV sizing by adjustment of balloon deployment volume avoided annular rupture in patients undergoing TAVR.

Для цитирования: Chavarria Jorge, Falcao Felipe, Dutra Gustavo, Sathananthan Janarthanan, Wood David, Douvi Georgia, Natarajan Madhu, Jaffer Iqbal, Velianou James, Sheth Tej. Bespoke valve sizing avoids annular rupture in patients treated with a balloon-expandable transcatheter heart valve.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2022. DOI: 10.1002/ccd.30402.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.