DOI 10.1016/j.jacc.2022.06.002
Standardized Definitions for Bioprosthetic Valve Dysfunction Following Aortic or Mitral Valve Replacement: JACC State-of-the-Art Review.
Аннотация
Bioprosthetic valve dysfunction (BVD) and bioprosthetic valve failure (BVF) may be caused by structural or nonstructural valve dysfunction. Both surgical and transcatheter bioprosthetic valves have limited durability because of structural valve deterioration. The main objective of this summary of experts participating in a virtual workshop was to propose standardized definitions for nonstructural and structural BVD and BVF following aortic or mitral biological valve replacement with the goal of facilitating research reporting and implementation of these terms in clinical practice. Definitions of structural BVF, based on valve reintervention or death, underestimate the true incidence of BVF. However, definitions solely based on the presence of high transprosthetic gradient at a given echocardiogram during follow-up overestimate the incidence of structural BVD and BVF. Definitions of aortic or mitral structural BVD must therefore include the confirmation by imaging of permanent structural changes to the leaflets alongside evidence of deterioration in valve hemodynamic function at echocardiography follow-up.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.