MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/j.jcrs.0000000000001428

Медицинская библиотека · Journal of cataract and refractive surgery · 2024

Clinical outcomes of a toric continuous range of vision presbyopia-correcting intraocular lens.

Orbegozo Javier, Pérez Angelica, Basterra Iñaki, Noguera Helena, González Laura, Lauzirika Gorka, Piñero David P

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the clinical outcomes of the toric version of a presbyopia-correcting intraocular lens (IOL) based on the combination of a diffractive-based extended depth-of-focus (EDOF) pattern and a diffractive multifocal platform.

SETTING

Miranza COI Bilbao, Bilbao, Spain.

DESIGN

Prospective case series.

METHODS

35 patients (51 to 84 years) with corneal astigmatism ranging from 0.75 to 2.19 diopters (D) undergoing bilateral cataract surgery with implantation of the Synergy Toric II IOL were evaluated during a 3-month follow-up. Visual acuity, refraction, defocus curve, and patient-reported outcomes with the Catquest-9SF questionnaire were analyzed. A vectorial analysis was used to analyze the accuracy of astigmatic correction.

RESULTS

Mean 3-month monocular postoperative uncorrected distance, intermediate (80 cm) and near (40 cm) visual acuities were 0.06 ± 0.11 logMAR, 0.13 ± 0.12 logMAR, and 0.13 ± 0.09 logMAR, respectively. Mean monocular distance-corrected intermediate (80 cm) and near visual acuity (40 cm) were 0.11 ± 0.12 logMAR and 0.10 ± 0.10 logMAR, respectively. Mean binocular defocus curve showed visual acuities of 0.10 logMAR or better for defocus levels from +0.50 to -2.50 D. Residual cylinder was within ±0.50 D in 97.0% of eyes. The surgically induced astigmatism prediction error ranged between -0.49 D and 0.50 D, with a mean value of 0.04 ± 0.16 D. Mean absolute IOL rotation was 3.79 ± 2.94 degrees. Significant improvements were found in all Rasch-calibrated scores obtained with Catquest-9SF ( P < .001).

CONCLUSIONS

The implantation of the toric presbyopia-correcting IOL evaluated provides an efficacious astigmatic correction while providing a fully restoration of the visual function across different distances.

Для цитирования: Orbegozo Javier, Pérez Angelica, Basterra Iñaki, Noguera Helena, González Laura, Lauzirika Gorka, Piñero David P. Clinical outcomes of a toric continuous range of vision presbyopia-correcting intraocular lens.. Journal of cataract and refractive surgery. 2024. DOI: 10.1097/j.jcrs.0000000000001428.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.