OBJECTIVE

Conventional otoscopes and oto-endoscopes, which are used to examine the tympanic membrane (TM), do not provide tomographic information. Optical coherence tomography (OCT) non-invasively reveals the depth-resolved internal microstructure of the TM with very high spatial resolution. We designed this study to examine the TMs with middle ear diseases using a handheld otoscope employing 860 nm spectral domain (SD)-OCT, combined with video camera and to demonstrate the clinical applicability of this system.

DESIGN

A total of 120 patients with otologic symptoms were enrolled. TM images were obtained using the handheld OCT-based otoscope (860 nm central wave length, 15 μm axial resolution, 15 μm lateral resolution, and 7 mm scanning range using relay lens). Both OCT and oto-endoscope images were compared according to the clinical characteristics such as perforation, retraction, and postoperative healing process.

RESULTS

The objective grade about the thickness of perforation margins and the accurate information about the extent of TM retraction that was not distinguishable by oto-endoscopic exam could be identified using this system. The postoperative healing process of TMs could be also followed using the OCT device.

CONCLUSION

These analyses from the surgeon-oriented perspective suggest another useful application of the handheld OCT device.

Для цитирования: Park Kibeom, Cho Nam Hyun, Jeon Mansik, Lee Sang Heun, Jang Jeong Hun, Boppart Stephen A, Jung Woonggyu, Kim Jeehyun. Optical assessment of the in vivo tympanic membrane status using a handheld optical coherence tomography-based otoscope.. Acta oto-laryngologica. 2017. DOI: 10.1080/00016489.2017.1395515.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.