DOI 10.1016/j.optom.2023.03.001

Медицинская библиотека · Journal of optometry · 2023

Accuracy and precision of automated subjective refraction in young hyperopes under cycloplegia.

Carracedo Gonzalo, Carpena-Torres Carlos, Pastrana Cristina, Rodríguez-Lafora María, Serramito María, Privado-Aroco Ana, Espinosa-Vidal Teresa María

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the agreement between the Eye Refract, an instrument to perform subjective automated refraction, and the traditional subjective refraction, as the gold standard, in young hyperopes under noncycloplegic and cycloplegic conditions.

METHODS

A cross-section and randomized study was carried out, involving 42 participants (18.2 ± 7.7 years, range 6 to 31 years). Only one eye was chosen for the analysis, randomly. An optometrist conducted the refraction with the Eye Refract, while another different optometrist conducted the traditional subjective refraction. Spherical equivalent (M), cylindrical components (J0 and J45), and corrected distance visual acuity (CDVA) were compared between both refraction methods under noncycloplegic and cycloplegic conditions. A Bland-Altman analysis was performed to assess the agreement (accuracy and precision) between both refraction methods.

RESULTS

Without cycloplegia, the Eye Refract showed significantly lower values of hyperopia than the traditional subjective refraction (p < 0.009), the mean difference (accuracy) and its 95% limits of agreement (precision) being -0.31 (+0.85, -1.47) D. Conversely, there were no statistical differences between both refraction methods under cycloplegic conditions (p ≥ 0.05). Regarding J0 and J45, both refraction methods manifested no significant differences between them under noncycloplegic and cycloplegic conditions (p ≥ 0.05). Finally, the Eye Refract significantly improved CDVA (0.04 ± 0.01 logMAR) compared with the traditional subjective refraction without cycloplegia (p = 0.01).

CONCLUSIONS

The Eye Refract is presented as a useful instrument to determine the refractive error in young hyperopes, the use of cycloplegia being necessary to obtain accurate and precise spherical refraction.

Для цитирования: Carracedo Gonzalo, Carpena-Torres Carlos, Pastrana Cristina, Rodríguez-Lafora María, Serramito María, Privado-Aroco Ana, Espinosa-Vidal Teresa María. Accuracy and precision of automated subjective refraction in young hyperopes under cycloplegia.. Journal of optometry. 2023. DOI: 10.1016/j.optom.2023.03.001.