Depth of focus of four novel extended range of vision intraocular lenses.

Schmid Ruediger, Fuchs Christin, Luedtke Holger, Borkenstein Andreas F

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the depth of focus of four latest non-diffractive extended range of vision IOLs on the optical bench. Such comparison had not been done before.

METHODS

We assessed and compared the through focus modulation transfer function (MTF) of the following novel IOLs with a nominal power of 22 D: Acrysof Vivity, LuxSmart Crystal, RayOne EMV and Tecnis Eyhance. An ISO-2 model eye was applied with apertures of 3 and 4.5 mm with monochromatic light of 546 nm. Measurements were done on OptiSpheric IOL PRO 2 optical bench.

RESULTS

For the aperture of 3 mm, Eyhance and RayOne EMV showed the most pronounced peak in MTF with only little enlarged depth of power. Vivity and LuxSmart showed two peaks of about 1.7 D respectively 1.3 D depth of focus, yet reduced MTF and with maxima differently located. For 4.5 mm, MTF values for Eyhance and particularly for RayOne EMV dropped. For Vivity and LuxSmart, only the peak for the secondary focus decreased.

CONCLUSION

Vivity and LuxSmart showed a larger depth of focus for our measuring conditions than Eyhance and RayOne EMV. Correspondingly, the peak MTF was best for Eyhance and RayOne ERV with small aperture. With the larger aperture, RayOne EMV considerably lost performance.

Для цитирования: Schmid Ruediger, Fuchs Christin, Luedtke Holger, Borkenstein Andreas F. Depth of focus of four novel extended range of vision intraocular lenses.. European journal of ophthalmology. 2022. DOI: 10.1177/11206721221125081.