Improved left ventricular unloading and circulatory support with synchronized pulsatile left ventricular assistance compared with continuous-flow left ventricular assistance in an acute porcine left ventricular failure model.

Letsou George V, Pate Thomas D, Gohean Jeffrey R, Kurusz Mark, Longoria Raul G, Kaiser Larry, Smalling Richard W

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Controversy exists regarding the optimal pumping method for left ventricular assist devices. The purpose of this investigation was to test the hypothesis that pulsatile left ventricular assist synchronized to the cardiac cycle provides superior left ventricular unloading and circulatory support compared with continuous-flow left ventricular assist devices at the same level of ventricular assist device flow.

METHODS

Seven male pigs were used to evaluate left ventricular assist device function using the TORVAD synchronized pulsatile-flow pump (Windmill Cardiovascular Systems, Inc, Austin, Tex) compared with the Bio-Medicus BPX-80 continuous-flow centrifugal pump (Medtronic, Inc, Minneapolis, Minn). Experiments were carried out under general anesthesia, and animals were instrumented via a median sternotomy. Hemodynamic measurements were obtained in the control state and with left ventricular assistance using the TORVAD and BPX-80 individually. Left ventricular failure was induced with suture ligation of the mid-left anterior descending coronary artery, and hemodynamic measurements were repeated.

RESULTS

During left ventricular assist device support, mean aortic pressure and total cardiac output were higher and left atrial pressure was lower with pulsatile compared with continuous flow at the same ventricular assist device flow rate. During ischemic left ventricular failure, pulsatile left ventricular support resulted in higher total cardiac output (5.58 ± 1.58 vs 5.12 ± 1.19, P < .05), higher mean aortic pressure (67.8 ± 14 vs 60.2 ± 10, P < .05), and lower left atrial pressure (11.5 ± 3.5 vs 13.9 ± 6.0, P < .05) compared with continuous flow at the same left ventricular assist device flow rate.

CONCLUSION

Synchronized, pulsatile left ventricular assistance produces superior left ventricular unloading and circulatory support compared with continuous-flow left ventricular assist at the same flow rates.

Для цитирования: Letsou George V, Pate Thomas D, Gohean Jeffrey R, Kurusz Mark, Longoria Raul G, Kaiser Larry, Smalling Richard W. Improved left ventricular unloading and circulatory support with synchronized pulsatile left ventricular assistance compared with continuous-flow left ventricular assistance in an acute porcine left ventricular failure model.. The Journal of thoracic and cardiovascular surgery. 2010. DOI: 10.1016/j.jtcvs.2010.03.043.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.