A randomized clinical trial to evaluate the safety and efficacy of a percutaneous left ventricular assist device versus intra-aortic balloon pumping for treatment of cardiogenic shock caused by myocardial infarction.

Seyfarth Melchior, Sibbing Dirk, Bauer Iris, Fröhlich Georg, Bott-Flügel Lorenz, Byrne Robert, Dirschinger Josef, Kastrati Adnan, Schömig Albert

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to test whether the left ventricular assist device (LVAD) Impella LP2.5 (Abiomed Europe GmbH, Aachen, Germany) provides superior hemodynamic support compared with the intra-aortic balloon pump (IABP).

BACKGROUND

Cardiogenic shock caused by left ventricular failure is associated with high mortality in patients with acute myocardial infarction (AMI). An LVAD may help to bridge patients to recovery from left ventricular failure.

METHODS

In a prospective, randomized study, 26 patients with cardiogenic shock were studied. The primary end point was the change of the cardiac index (CI) from baseline to 30 min after implantation. Secondary end points included lactic acidosis, hemolysis, and mortality after 30 days.

RESULTS

In 25 patients the allocated device (n = 13 IABP, n = 12 Impella LP2.5) could be safely placed. One patient died before implantation. The CI after 30 min of support was significantly increased in patients with the Impella LP2.5 compared with patients with IABP (Impella: DeltaCI = 0.49 +/- 0.46 l/min/m(2); IABP: DeltaCI = 0.11 +/- 0.31 l/min/m(2); p = 0.02). Overall 30-day mortality was 46% in both groups.

CONCLUSIONS

In patients presenting with cardiogenic shock caused by AMI, the use of a percutaneously placed LVAD (Impella LP 2.5) is feasible and safe, and provides superior hemodynamic support compared with standard treatment using an intra-aortic balloon pump. (Efficacy Study of LV Assist Device to Treat Patients With Cardiogenic Shock [ISAR-SHOCK]; NCT00417378).

Для цитирования: Seyfarth Melchior, Sibbing Dirk, Bauer Iris, Fröhlich Georg, Bott-Flügel Lorenz, Byrne Robert, Dirschinger Josef, Kastrati Adnan, Schömig Albert. A randomized clinical trial to evaluate the safety and efficacy of a percutaneous left ventricular assist device versus intra-aortic balloon pumping for treatment of cardiogenic shock caused by myocardial infarction.. Journal of the American College of Cardiology. 2008. DOI: 10.1016/j.jacc.2008.05.065.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.