DOI 10.1161/circulationaha.106.664672
BioSTAR Evaluation STudy (BEST): a prospective, multicenter, phase I clinical trial to evaluate the feasibility, efficacy, and safety of the BioSTAR bioabsorbable septal repair implant for the closure of atrial-level shunts.
Аннотация
BACKGROUND
The use of permanent synthetic implants to close atrial septal defects (ASD) and patent foramen ovale (PFO) has a number of limitations, including late complications and the limiting of transeptal access to the left heart should it be required for the later treatment of acquired heart disease. BioSTAR is a novel, bioabsorbable, atrial septal repair implant. This phase I pilot study evaluates the feasibility, safety, and effectiveness of BioSTAR for the first time in humans.
METHODS AND
RESULTS
We conducted a prospective, open-label, multicenter clinical study in 58 patients aged 28 to 68 years who had a clinically significant ASD or PFO. Percutaneous shunt closure was undertaken with the BioSTAR septal repair implant. Successful device implantation was achieved in 57 (98%) of 58 patients. Closure at 30 days and 6 months, assessed by contrast transthoracic echocardiography, was 48 (92%) of 52 and 54 (96%) of 56, respectively. There was no evidence of a clinically significant response to the device. Transient atrial arrhythmia occurred in 5 patients after implantation. No major safety issues were observed.
CONCLUSIONS
This study demonstrates the feasibility, safety, and effectiveness of BioSTAR for the closure of ASD and PFO in humans with a high rate of early and complete shunt closure. BioSTAR is a novel septal repair implant designed to provide biological closure of atrial-level defects using the patient's natural healing response. Because 90% to 95% of the implant is absorbed and replaced with healthy native tissue, future access to the left atrium may be achieved.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.