MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/00016489.2017.1354395

Медицинская библиотека · Acta oto-laryngologica · 2017

Comparative study between the caloric vestibular and the video-head impulse tests in unilateral Menière's disease.

Cordero-Yanza Jonathan Andrés, Arrieta Vázquez Edna Verónica, Hernaiz Leonardo Juan Carlos, Mancera Sánchez Jimena, Hernández Palestina Mario Sabas, Pérez-Fernández Nicolás

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Compare the caloric vestibular test (CVT) and the video head impulse test (vHIT) in the evaluation of Meniere's disease (MD) and to analyze their diagnostic accuracy.

MATERIALS AND

METHODS

Retrospective review of CVT and vHIT performed in MD patients at a tertiary care referral center in Spain.

RESULTS

A total of 88 patients were reviewed. The CVT was abnormal in 67% (n = 59) of patients, while the vHIT was abnormal in 66% (n = 58) of them. Agreement between both tests was poor, regardless of whether the horizontal semicircular canal (SCC) or all SCC were analyzed (kappa = .21). Anterior SCC pathology was found in 30% (n = 27) of the subjects while 51% (n = 45) had altered gains in the posterior SCC.

CONCLUSIONS

Our study shows that patients with MD can have functional deficit involving the vertical semicircular canals, which cannot be detected by the CVT alone. Therefore, both tests should be used in a complementary fashion.

Для цитирования: Cordero-Yanza Jonathan Andrés, Arrieta Vázquez Edna Verónica, Hernaiz Leonardo Juan Carlos, Mancera Sánchez Jimena, Hernández Palestina Mario Sabas, Pérez-Fernández Nicolás. Comparative study between the caloric vestibular and the video-head impulse tests in unilateral Menière's disease.. Acta oto-laryngologica. 2017. DOI: 10.1080/00016489.2017.1354395.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.