MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s12028-010-9459-6

Медицинская библиотека · Neurocritical care · 2011

Lumbar catheter for monitoring of intracranial pressure in patients with post-hemorrhagic communicating hydrocephalus.

Speck Verena, Staykov Dimitre, Huttner Hagen B, Sauer Roland, Schwab Stefan, Bardutzky Juergen

Клиническое исследованиеСтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

We investigated the feasibility and accuracy of intracranial pressure (ICP)-measurement by lumbar drainage (LD) catheter in patients with post-hemorrhagic communicating hydrocephalus (PHCH).

METHODS

Patients with subarachnoid hemorrhage (SAH, n = 21) or spontaneous ganglionic hemorrhage (ICH, n = 22) with ventricular involvement and the need for external ventricular drainage (EVD) due to acute hydrocephalus were included. When EVD weaning was not feasible due to persistent hydrocephalus, an additional LD was placed, after which EVD was clamped off. During this overlap period, patients underwent simultaneous pressure recording via EVD ("EVD-ICP") and LD ("LD-ICP"). Testing included manual compression of the jugular veins and body-posture changes from supine to 30° position. After EVD removal, we evaluated sensitivity and specificity of ICP-rise >20 mmHg during continuous monitoring via LD for the detection of persistent PHCH using additional evaluation with computed tomography (CT).

RESULTS

A total of 1,806 measurements were performed in 43 patients. "LD-ICP" was strongly correlated to "EVD-ICP", with determination coefficients R(2) for the baseline measurements and each of the maneuvers ranging from 0.95-0.99, and slopes ranging 0.96-1.01. Sensitivity of "LD-ICP" >20 mmHg for detection of persistent PHCH as compared to CT was 81% and specificity was 100%. Two patients with severe SAH developed reversible signs of herniation after gradually increasing differences between "LD-ICP" and "EVD-ICP" indicated a cranio-spinal pressure gradient, likely due to cerebrospinal fluid overdrainage via LD.

CONCLUSION

ICP measured via LD highly and reliably correlated to ICP measured via EVD in patients with PHCH.

Для цитирования: Speck Verena, Staykov Dimitre, Huttner Hagen B, Sauer Roland, Schwab Stefan, Bardutzky Juergen. Lumbar catheter for monitoring of intracranial pressure in patients with post-hemorrhagic communicating hydrocephalus.. Neurocritical care. 2011. DOI: 10.1007/s12028-010-9459-6.