Effect of Intraocular Lens Power Calculation Formula Optimization in the Sum-of-Segments Optical Biometer.

Kojima Takashi, Tamaoki Akeno, Ichikawa Kazuo, Satoh Yuya, Tomemori Ryota, Watanabe Keizo, Hasegawa Asato, Sawaki Ayako, Kaga Tatsushi

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: We evaluated the effect of optimization of the intraocular lens (IOL) power calculation formula SRK/T and Barrett Universal II (BU II) in long eyes (≥26 mm: group L) and short eyes (≤22 mm: group S) using axial length calculated from segmented refractive indices (SRI).

SETTING

Multicenter study at five sites in Japan.

DESIGN

Retrospective observational study.

METHODS

This study included 461 eyes of 461 patients (mean age 73.8 ± 8.4 years) who underwent cataract surgery. The predicted refractive error (PRE) was compared between the SRI (ARGOS) and the equivalent refractive index (ERI) biometers (IOLMasterTM700). The patients were randomly divided into two groups, a learning group and a validation group. The optimization constants were determined in the learning group, and the optimization constants were subsequently applied to the validation group and compared with the ERI biometer results.

RESULTS

Using both SRK/T and BU II, the validation group's PRE using optimization constants for the SRI biometer in group L was significantly smaller than that using the ERI biometer (p<0.001, p<0.01). In group L, the arithmetic PRE of Barrett UII formula with SRI showed a significant improvement after optimization compared to before optimization (p<0.0001). In group S, the arithmetic PRE of SRK/T and Barrett UII formula with SRI showed a significant improvement (p<0.0001, p<0.0001).

CONCLUSION

In long and short eyes, the current study revealed that optimization of the SRK/T and Barrett formula constants for the SRI biometer was beneficial to achieve accurate refractive outcomes after cataract surgery.

Для цитирования: Kojima Takashi, Tamaoki Akeno, Ichikawa Kazuo, Satoh Yuya, Tomemori Ryota, Watanabe Keizo, Hasegawa Asato, Sawaki Ayako, Kaga Tatsushi. Effect of Intraocular Lens Power Calculation Formula Optimization in the Sum-of-Segments Optical Biometer.. Clinical ophthalmology (Auckland, N.Z.). 2024. DOI: 10.2147/opth.s477006.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.