Journal of cataract and refractive surgery · 2024
Spherical equivalent prediction analysis in intraocular lens power calculations using Eyetemis: a comprehensive approach.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Kan-Tor Yoav, Abulafia Adi, Zadok David, Kohnen Thomas, Savini Giacomo, Hoffer Kenneth J, Benjamini Yuval
Аннотация
PURPOSE: To compare 2 different datasets, using Eyetemis, an online analytical tool designed for assessing the spherical equivalent prediction errors (SEQ-PEs) of intraocular lens (IOL) power calculation formulas after cataract surgery.
SETTING: Institutional.
DESIGN: Retrospective case series.
METHODS: The study comprised 2 distinct datasets of patients who had undergone successful cataract surgery. Dataset 1 includes standard eyes, whereas Dataset 2 includes eyes with keratoconus. An online tool was used for SEQ-PE analysis across the 2 datasets, adhering to ISO standards for evaluating accuracy based on trueness and precision. The tool incorporates robust t tests for comparing the trimmed mean of the data, adjusting for heteroscedasticity. IOL constants in Dataset 1 were optimized for the comparison of Hoffer Q, Holladay 1, SRK/T, Haigis, and Barrett Universal II (BUII) formulas. In Dataset 2, IOL constants from the IOLCon website were used for the comparison of the BUII and its designated KCN version: Barrett TrueK Keratoconus (TrueK [KCN]).
RESULTS: For Dataset 1, the trimmed mean SEQ-PE values of all formulas were not significantly different from zero. BUII had superior precision and accuracy compared with all other formulas, except from Haigis ( P ≤ .04). For Dataset 2, BUII's trimmed-mean SEQ-PE was significantly different from zero (0.59 diopters [D], P < .01), unlike the TrueK (KCN) (0.12 D, P = .10). In addition, TrueK (KCN) exhibited enhanced precision and accuracy relative to BUII ( P < .01).
CONCLUSIONS: The online analysis tool provides a streamlined approach for assessing the prediction accuracy of SEQ refraction after cataract surgery, effectively evaluating trueness, precision, and overall accuracy through the use of advanced statistical methods.
SETTING: Institutional.
DESIGN: Retrospective case series.
METHODS: The study comprised 2 distinct datasets of patients who had undergone successful cataract surgery. Dataset 1 includes standard eyes, whereas Dataset 2 includes eyes with keratoconus. An online tool was used for SEQ-PE analysis across the 2 datasets, adhering to ISO standards for evaluating accuracy based on trueness and precision. The tool incorporates robust t tests for comparing the trimmed mean of the data, adjusting for heteroscedasticity. IOL constants in Dataset 1 were optimized for the comparison of Hoffer Q, Holladay 1, SRK/T, Haigis, and Barrett Universal II (BUII) formulas. In Dataset 2, IOL constants from the IOLCon website were used for the comparison of the BUII and its designated KCN version: Barrett TrueK Keratoconus (TrueK [KCN]).
RESULTS: For Dataset 1, the trimmed mean SEQ-PE values of all formulas were not significantly different from zero. BUII had superior precision and accuracy compared with all other formulas, except from Haigis ( P ≤ .04). For Dataset 2, BUII's trimmed-mean SEQ-PE was significantly different from zero (0.59 diopters [D], P < .01), unlike the TrueK (KCN) (0.12 D, P = .10). In addition, TrueK (KCN) exhibited enhanced precision and accuracy relative to BUII ( P < .01).
CONCLUSIONS: The online analysis tool provides a streamlined approach for assessing the prediction accuracy of SEQ refraction after cataract surgery, effectively evaluating trueness, precision, and overall accuracy through the use of advanced statistical methods.