Journal francais d'ophtalmologie · 2024
Comparison of refractive outcomes obtained with two swept-source OCT-based optical biometers after cataract surgery: A study of 152 eyes.
СтатьяСравнительное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Agard E, Levron A, Billant J, Douma I, Dot C
Аннотация
PURPOSE: To compare two swept-source OCT optical biometers, the Anterion® (Heidelberg Engineering GmbH, Heidelberg, Germany) and the IOLMaster 700® (Carl Zeiss Meditec AG, Jena, Germany), in the analysis of biometric data, intraocular lens (IOL) calculation and postoperative spherical equivalent predictability.
METHODS: This was a real-life, single-center, retrospective study including 152 eyes of 81 consecutive patients referred for cataract surgery. All patients were examined with the IOLMaster 700® and the Anterion®. Biometric data (axial length [AL], anterior chamber depth [ACD], mean keratometry [Km], K1 [flat keratometry], K2 [steep keratometry] and axis, TK1 [flat total keratometry], TK2 [steep total keratometry], central pachymetry, lens thickness [LT], white-to-white distance [WTW]), IOL calculation with the SRK/T formula, and postoperative refractive outcome at 1 month were compared.
RESULTS: All biometric measurements were significantly different between the two biometers. Correlations were excellent for AL, pachymetry, ACD, LT and keratometry measurements, and for the IOL calculation (r>0.96, intraclass correlation coefficient=1). The IOL power for emmetropia was similar between both biometers when the SRK/T formula was used (20.84±3.24D versus 20.86±3.29D, P=0.61). The mean postoperative spherical equivalent prediction error calculated using the SRK/T formula was 0.03±0.5D for the IOLMaster 700® versus 0.01±0.47D for the Anterion® (P=0.12).
CONCLUSIONS: This study showed excellent correlation and agreement for the biometric measurements and the IOL power calculation with the SRK/T formula between both biometers.
METHODS: This was a real-life, single-center, retrospective study including 152 eyes of 81 consecutive patients referred for cataract surgery. All patients were examined with the IOLMaster 700® and the Anterion®. Biometric data (axial length [AL], anterior chamber depth [ACD], mean keratometry [Km], K1 [flat keratometry], K2 [steep keratometry] and axis, TK1 [flat total keratometry], TK2 [steep total keratometry], central pachymetry, lens thickness [LT], white-to-white distance [WTW]), IOL calculation with the SRK/T formula, and postoperative refractive outcome at 1 month were compared.
RESULTS: All biometric measurements were significantly different between the two biometers. Correlations were excellent for AL, pachymetry, ACD, LT and keratometry measurements, and for the IOL calculation (r>0.96, intraclass correlation coefficient=1). The IOL power for emmetropia was similar between both biometers when the SRK/T formula was used (20.84±3.24D versus 20.86±3.29D, P=0.61). The mean postoperative spherical equivalent prediction error calculated using the SRK/T formula was 0.03±0.5D for the IOLMaster 700® versus 0.01±0.47D for the Anterion® (P=0.12).
CONCLUSIONS: This study showed excellent correlation and agreement for the biometric measurements and the IOL power calculation with the SRK/T formula between both biometers.