DOI 10.1016/j.jfo.2024.104186

Медицинская библиотека · Journal francais d'ophtalmologie · 2024

Comparison of refractive outcomes obtained with two swept-source OCT-based optical biometers after cataract surgery: A study of 152 eyes.

Agard E, Levron A, Billant J, Douma I, Dot C

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare two swept-source OCT optical biometers, the Anterion® (Heidelberg Engineering GmbH, Heidelberg, Germany) and the IOLMaster 700® (Carl Zeiss Meditec AG, Jena, Germany), in the analysis of biometric data, intraocular lens (IOL) calculation and postoperative spherical equivalent predictability.

METHODS

This was a real-life, single-center, retrospective study including 152 eyes of 81 consecutive patients referred for cataract surgery. All patients were examined with the IOLMaster 700® and the Anterion®. Biometric data (axial length [AL], anterior chamber depth [ACD], mean keratometry [Km], K1 [flat keratometry], K2 [steep keratometry] and axis, TK1 [flat total keratometry], TK2 [steep total keratometry], central pachymetry, lens thickness [LT], white-to-white distance [WTW]), IOL calculation with the SRK/T formula, and postoperative refractive outcome at 1 month were compared.

RESULTS

All biometric measurements were significantly different between the two biometers. Correlations were excellent for AL, pachymetry, ACD, LT and keratometry measurements, and for the IOL calculation (r>0.96, intraclass correlation coefficient=1). The IOL power for emmetropia was similar between both biometers when the SRK/T formula was used (20.84±3.24D versus 20.86±3.29D, P=0.61). The mean postoperative spherical equivalent prediction error calculated using the SRK/T formula was 0.03±0.5D for the IOLMaster 700® versus 0.01±0.47D for the Anterion® (P=0.12).

CONCLUSIONS

This study showed excellent correlation and agreement for the biometric measurements and the IOL power calculation with the SRK/T formula between both biometers.

Для цитирования: Agard E, Levron A, Billant J, Douma I, Dot C. Comparison of refractive outcomes obtained with two swept-source OCT-based optical biometers after cataract surgery: A study of 152 eyes.. Journal francais d'ophtalmologie. 2024. DOI: 10.1016/j.jfo.2024.104186.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.