Association of refractive outcome with postoperative anterior chamber depth measured with 3 optical biometers.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mao Yan, Li Jianbing, Qin Yingyan, Xu Yanxin, Liu Liangping, Cheng Huanhuan, Wu Mingxing
Аннотация
PURPOSE: This study evaluated the relationship between refractive outcomes and postoperative anterior chamber depth (ACD, measured from corneal epithelium to lens) measured by swept-source optical coherence tomography (SS-OCT), optical low-coherence reflectometry (OLCR), and Scheimpflug devices under the undilated pupil.
METHODS: Patients undergoing cataract phacoemulsification with intraocular lens (IOL) implantation in a hospital setting were enrolled. Postoperative ACD (postACD) was performed with an SS-OCT device, an OLCR device, and a Scheimpflug device at least 1 month after cataract surgery. After adjusting the mean predicted error to 0, differences in refractive outcomes were calculated with the Olsen formula using actual postACD measured from 3 devices and predicted value.
RESULTS: Overall, this comparative case study included 69 eyes of 69 patients, and postACD measurements were successfully taken using all 3 devices. The postACD measured with the SS-OCT, OLCR, and Scheimpflug devices was 4.59 ± 0.30, 4.50 ± 0.30, and 4.54 ± 0.32 mm, respectively. Statistically significant differences in postACD were found among 3 devices (P < 0.001), with intraclass correlation coefficients (ICCs) and Bland-Altman showing good agreement. No significant difference in median absolute error was found with the Olsen formula using actual postACD obtained with 3 devices. Percentage prediction errors were within ± 0.50 D in 65% (OLCR), 70% (Scheimpflug), and 67% (SS-OCT) calculated by actual postACD versus 64% by predicted value.
CONCLUSION: Substantial agreement was found in postACD measurements obtained from the SS-OCT, OLCR, and Scheimpflug devices, with a trend toward comparable refractive outcomes in the Olsen formula. Meanwhile, postACD measurements may be potentially superior for the additional enhancement of refractive outcomes.
METHODS: Patients undergoing cataract phacoemulsification with intraocular lens (IOL) implantation in a hospital setting were enrolled. Postoperative ACD (postACD) was performed with an SS-OCT device, an OLCR device, and a Scheimpflug device at least 1 month after cataract surgery. After adjusting the mean predicted error to 0, differences in refractive outcomes were calculated with the Olsen formula using actual postACD measured from 3 devices and predicted value.
RESULTS: Overall, this comparative case study included 69 eyes of 69 patients, and postACD measurements were successfully taken using all 3 devices. The postACD measured with the SS-OCT, OLCR, and Scheimpflug devices was 4.59 ± 0.30, 4.50 ± 0.30, and 4.54 ± 0.32 mm, respectively. Statistically significant differences in postACD were found among 3 devices (P < 0.001), with intraclass correlation coefficients (ICCs) and Bland-Altman showing good agreement. No significant difference in median absolute error was found with the Olsen formula using actual postACD obtained with 3 devices. Percentage prediction errors were within ± 0.50 D in 65% (OLCR), 70% (Scheimpflug), and 67% (SS-OCT) calculated by actual postACD versus 64% by predicted value.
CONCLUSION: Substantial agreement was found in postACD measurements obtained from the SS-OCT, OLCR, and Scheimpflug devices, with a trend toward comparable refractive outcomes in the Olsen formula. Meanwhile, postACD measurements may be potentially superior for the additional enhancement of refractive outcomes.