Rotational alignment of the lower extremity in the presence of total knee endoprosthesis: Reproducibility of torsion analyses using ultra-low-dose photon-counting CT.

Patzer Theresa Sophie, Kunz Andreas Steven, Huflage Henner, Luetkens Karsten Sebastian, Conrads Nora, Pannenbecker Pauline, Jakuscheit Axel, Reppenhagen Stephan, Ergün Süleyman, Bley Thorsten Alexander, Grunz Jan-Peter

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Leg torsion analysis can provide valuable information in symptomatic patients after total knee arthroplasty. However, extensive beam-hardening and photon-starvation artifacts limit diagnostic assessability and dose reduction potential. For this study, we investigated the reproducibility of rotational measurements in ultra-low-dose photon-counting CT with spectral shaping via tin prefiltration.

MATERIAL AND

METHODS

Employing a first-generation photon-counting CT, eight cadaveric specimens were examined with an established three-level scan protocol (hip: Sn 140, knee: Sn 100, ankle: Sn 100 kVp). In three body donors with unilateral knee endoprostheses, additional modified settings were applied (Sn 140 kVp at knee level). Protocols were executed with three dose levels (hip-knee-ankle, high-quality: 5.0-3.0-2.0 mGy, low-dose: 0.80-0.30-0.26 mGy, ultra-low-dose: 0.25-0.06-0.06 mGy). Six radiologists performed torsion analyses, additionally reporting their diagnostic confidence. Intraclass correlation coefficients (ICC) were calculated to assess interrater reliability.

RESULTS

No significant differences were ascertained for femoral (p = 0.330), tibial (p = 0.177), and overall leg rotation measurements (p = 0.358) among high-quality, low-dose, and ultra-low-dose protocols. Interrater reliability was excellent for torsion of the femur (ICC 0.915, 95% confidence interval 0.871-0.947), tibia (0.960, 0.938-0.976), and overall leg (0.967, 0.945-0.981). In specimens with total knee endoprostheses, absolute rotational measurements were unaffected by dose level and tube voltage despite superior diagnostic confidence on the ipsilateral and contralateral sides with modified settings (p < 0.001).

CONCLUSIONS

Combining the advantages of photon-counting CT and spectral shaping, reliable leg torsion analyses are feasible with ultra-low radiation exposure even in the presence of total knee endoprostheses.

Для цитирования: Patzer Theresa Sophie, Kunz Andreas Steven, Huflage Henner, Luetkens Karsten Sebastian, Conrads Nora, Pannenbecker Pauline, Jakuscheit Axel, Reppenhagen Stephan, Ergün Süleyman, Bley Thorsten Alexander, Grunz Jan-Peter. Rotational alignment of the lower extremity in the presence of total knee endoprosthesis: Reproducibility of torsion analyses using ultra-low-dose photon-counting CT.. European journal of radiology. 2023. DOI: 10.1016/j.ejrad.2023.111055.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.