Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2020

Active Transcutaneous Bone Conduction Implant: Audiometric Outcomes Following a Novel Middle Fossa Approach With Self-Drilling Screws.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Siegel Lauren, You Peng, Zimmerman Kim, Parnes Lorne, Agrawal Sumit K

Аннотация

OBJECTIVE: To present surgical and audiometric outcomes of patients implanted with an active transcutaneous bone conduction implant following the novel middle fossa surgical approach with self-drilling screws.

STUDY DESIGN: Retrospective review.

SETTING: Tertiary care center.

PATIENTS: Thirty-seven adults with either conductive or mixed hearing loss that met indications for an active transcutaneous bone conduction implant were consecutively implanted from April, 2013 to May, 2018.

INTERVENTION: Unilateral middle fossa implantation of an active transcutaneous bone conduction implant.

MAIN OUTCOME MEASURES: Patient charts were reviewed for surgical outcomes and complications over the 6-year period. Preoperative air conduction, preoperative bone conduction, and 3-month postoperative aided thresholds were recorded. Speech perception was assessed using CNC words and AzBio sentences. Pure-tone averages (PTAs; measured at 0.5, 1.0, 2.0 and 3.0 kHz), air-bone gap, and functional gain were calculated.

RESULTS: Mean air conduction and bone conduction PTAs (±standard deviation) of the implanted ear were 66.8 dB (±14.9 dB) and 21.9 dB (±14.0 dB), respectively. Mean aided PTA was 26.5 dB (± 8.5 dB). The average functional gain was 40.3 dB (±19.0 dB). Favorable speech perception outcomes were observed. No complications or instances of revision surgery were reported, with a mean follow-up time of 32 months (range, 9-71 mo).

CONCLUSIONS: This is the first paper to describe outcomes of patients implanted with an active transcutaneous bone conduction implant via the middle fossa with self-drilling screws. Favorable surgical outcomes were observed with a follow-up of up to 6 years.