Preloaded vs manually loaded intraocular lens delivery failure modes: an analysis of Food and Drug Administration device reports.

Hecht Idan, Tuuminen Raimo, Kanclerz Piotr, Sella Ruti

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare reported intraoperative failure-mode profiles between preloaded and manually loaded intraocular lens (IOL) delivery systems using U.S. Food and Drug Administration (FDA) device reports.

SETTING

FDA Manufacturer and User Facility Device Experience (MAUDE) database, United States.

DESIGN

Retrospective cross-sectional study.

METHODS

MAUDE reports from January 1, 2016 through December 31, 2025 were reviewed. Reports describing intraoperative IOL loading, delivery, or positioning issues were included when injector type was specified. Structured MAUDE fields served as the primary data source, and free-text narratives were analyzed using automated natural language processing to supplement classification when consistent with structured fields. Comparisons were performed using within-group proportions, and multivariable logistic regression estimated adjusted odds ratios (aORs) controlling for report year, manufacturer, lens platform, and reporting entity.

RESULTS

Of 57,435 FDA device reports, 14,460 intraoperative reports met inclusion criteria; 10,567 (73.1%) involved preloaded systems and 3,893 (26.9%) manually loaded systems. Manually loaded systems showed higher reporting of IOL damage (66.0% vs 45.4%; aOR 3.06, 95% CI 2.78-3.37; p<0.001) and loading issues (aOR 5.53, 95% CI 3.99-7.66; p<0.001). Preloaded systems showed higher reporting of delivery failures (40.5% vs 21.8%; aOR 3.91, 95% CI 3.51-4.35; p<0.001), lens positioning issues (aOR 1.27, 95% CI 1.01-1.62; p=0.041), and material or optics abnormalities (aOR 3.53, 95% CI 1.96-6.35; p<0.001).

CONCLUSIONS

Preloaded and manually loaded IOL delivery systems demonstrated distinct failure-mode profiles within FDA device reports. These findings characterized reporting patterns rather than population-level risk and may inform device design and post-market surveillance priorities.

Для цитирования: Hecht Idan, Tuuminen Raimo, Kanclerz Piotr, Sella Ruti. Preloaded vs manually loaded intraocular lens delivery failure modes: an analysis of Food and Drug Administration device reports.. Journal of cataract and refractive surgery. 2026. DOI: 10.1097/j.jcrs.0000000000001977.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.