Iridociliary measurements using the Anterion swept-source optical coherence tomography.
Аннотация
PURPOSE: Anterior segment optical coherence tomography (AS-OCT) by a swept source can visualize the ciliary body. The study was performed for analyzing the feasibility of a new swept-source OCT (Anterion) device for measuring iris ciliary sulcus, ciliary body thickness, and iris thickness.
METHODS
It is a pilot study among 30 normal young participants and hospital employees with their consent. The ciliary body, iris thickness, and ciliary sulcus were measured and compared to the results in the literature. We obtained 28 good-quality results. The main outcome measures were iris thickness (near and away from the pupillary margin), ciliary body thickness, and iridociliary sulcus measurement.
RESULTS
The iris thickness 2.5 mm from the pupillary margin was 0.70 ± 0.10 mm in the right eye and 0.68 ± 0.11 in the left eye. Closer to the root, the iris thickness varied from 0.55 ± 0.16 mm (right eye) and 0.57 ± 0.12 (left eye). Ciliary body thickness: Right eye 0.59 ± 0.14mm (28 good images), left eye 0.58 ± 0.13mm; ciliary sulcus (inner angle): right eye 82.65 ± 26.6°, left eye 83.66 ± 30.1°.
CONCLUSION
Anterion OCT is feasible in our setting for various measurements and can provide valuable inputs to ophthalmologists for the treatment of ocular diseases.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.