PURPOSE: The aortic root pressurizing device was developed for aortic valve repair surgery. It allows echocardiographic evaluation of the aortic valve during cardioplegic arrest by mimicking diastolic afterload.

DESCRIPTION: This polyoxymethylene or polyether ether ketone-based device consists of a sealing cap nut, a sealing ring, a plug screw, and both a filling and a ventilation line. It can be easily connected to any size of aortic Dacron grafts.

EVALUATION: The device was tested in 15 porcine hearts using a beating heart biosimulator including hemodynamic and echocardiographic monitoring. Valve competence was analyzed on both the beating and resting heart at 60 and 45 mm Hg root pressure. Aortic insufficiency was surgically induced by leaflet manipulation. Native aortic valves showed either none or trivial aortic insufficiency. After leaflet manipulation echocardiographic proof of valve insufficiency was possible in all specimen. Jet direction was identic in all cases at 60 mm Hg and 93% at 45 mm Hg root pressure.

CONCLUSIONS

The aortic root pressurizing device shows highly comparable results of echocardiographic aortic valve evaluation between static and dynamic settings under experimental conditions.

Для цитирования: Steindl Johannes, Kirnbauer Michael, Fierlbeck Johann, Krieghofer Michael, Hitzl Wolfgang, Hinterbuchner Lynne, Seitelberger Rainald, Dinges Christian. Aortic Root Pressurizing Device: Aortic Valve Evaluation During Cardioplegic Arrest.. The Annals of thoracic surgery. 2019. DOI: 10.1016/j.athoracsur.2019.10.080.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.