Intraoperative optical coherence tomography measurements of aphakic eyes to predict postoperative position of 2 intraocular lens designs.

Hirnschall Nino, Farrokhi Sanaz, Amir-Asgari Sahand, Hienert Julius, Findl Oliver

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate intraoperative anterior chamber depth (ACD) measurements of the aphakic eye to predict the postoperative ACD and compare 2 intraocular lens (IOL) designs.

SETTING

Hanusch Hospital, Vienna, Austria.

DESIGN

Prospective study.

METHODS

In this prospective study, patients scheduled for cataract surgery received a plate-haptic IOL (Asphina, Carl Zeiss Meditec AG) or an open-loop haptic IOL (ZCB00, Johnson & Johnson). Preoperatively, optical biometry (IOLMaster 700, CZM, or Lenstar, Haag-Streit) was performed. Intraoperatively, a prototype setup was used to perform time-domain OCT scans of the anterior eye segment (Visante connected to OPMI Lumera 200, both CZM). The intraoperative ACD was measured and used to predict the postoperative IOL position. Optical biometry and subjective refraction and autorefraction (RM 8800, Topcon) were performed 2 months postoperatively.

RESULTS

The study comprised 203 eyes of 203 patients. A partial least-square regression model for ACD generated 2 months postoperatively showed that the predictive power of the intraoperative ACD (0.48) was highest followed by the axial eye length (0.45) and then the preoperatively measured ACD (0.30). These findings were confirmed in a bootstrapping model. Regression models combining the preoperative ACD and intraoperative ACD resulted in further significant improvement.

CONCLUSIONS

Intraoperative ACD measurements predicted the postoperative position of open-loop IOLs and plate-haptic IOLs better than preoperative ACD measurements. Combining preoperative and intraoperative ACD measurements further improved the prediction.

Для цитирования: Hirnschall Nino, Farrokhi Sanaz, Amir-Asgari Sahand, Hienert Julius, Findl Oliver. Intraoperative optical coherence tomography measurements of aphakic eyes to predict postoperative position of 2 intraocular lens designs.. Journal of cataract and refractive surgery. 2018. DOI: 10.1016/j.jcrs.2018.07.044.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.