SILICONE MICROTUBE-ASSISTED SCLERAL FIXATION OF A POSTERIOR CHAMBER INTRAOCULAR LENS.

Kataoka Takuya, Kamei Motohiro

СтатьяVideo-Audio MediaПеревод пока не подготовлен

Аннотация

PURPOSE: To report a novel technique for intrascleral fixation of a posterior chamber intraocular lens using a silicone microtube to manipulate the haptics into position.

METHODS

Intrascleral fixation was performed in six eyes and the results evaluated in this retrospective case series. A silicone microtube with a 0.2-mm external diameter was passed from a sclerocorneal incision through the chamber and a sclerotomy made using a 30-gauge needle. The tips of the intraocular lens haptics were connected to the silicone microtube outside the eye. After the intraocular lens was injected into the posterior chamber, the haptics were drawn through the scleral incision through their attachment to the silicone microtube.

RESULTS

The mean postoperative corrected visual acuity was 0.62 logarithm of the minimum angle of resolution (20/43) with a mean refraction error of -0.06 ± 0.4 diopter, which did not differ significantly (P = 0.53) from the expected value. The postoperative complications included transient ocular hypotension, vitreous hemorrhage, and choroidal detachment.

CONCLUSION

Our technique using a silicone microtube reduces the number of intraocular procedures compared with previous methods using forceps or needles for moving the intraocular lens haptics from the posterior chamber to the outside through sclerotomies.

Для цитирования: Kataoka Takuya, Kamei Motohiro. SILICONE MICROTUBE-ASSISTED SCLERAL FIXATION OF A POSTERIOR CHAMBER INTRAOCULAR LENS.. Retina (Philadelphia, Pa.). 2018. DOI: 10.1097/iae.0000000000002143.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.