Accuracy of formulae for secondary intraocular lens power calculations in pediatric aphakia.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Nakhli Fouad R, Emarah Khaled, Jeddawi Laila
Аннотация
PURPOSE: To compare the accuracy of axial length vergence formulas versus refractive vergence formulas for secondary intraocular lens (IOL) implantation in pediatric aphakia.
METHODS: This retrospective comparative study, evaluated 31 eyes of 31 patients aged ≤3.5 years, who had undergone secondary IOL implantation. The median absolute error (MedAE) was compared between axial length vergence formulas (Hoffer Q, Holladay I, SRK II, and SRK/T) and refractive vergence formulas (Lanchulev, Holladay R, Mackool, and Khan) as well as between formulas within the same vergence.
RESULTS: There was a significant difference (P = 0.010) between MedAE for axial length vergence formulas [1.19 Diopter(D)] and MedAE for refractive vergence formulas (2.48 D). The MedAE of axial length vergence formulas were comparable as to Hoffer (1.59 D), Holladay (1.27 D), SRK/T (1.23 D), and SRK II (1.30 D). Among refractive vergence formulas, Lanchulev (5.00 D) and Holladay R (2.51 D) had significantly larger MedAE as compared to Khan (2.06 D) and Mackool (2.15 D).
CONCLUSION: Axial length vergence formulas performed significantly better than refractive vergence formulas; however, axial length vergence formulas were comparable within the same vergence.
METHODS: This retrospective comparative study, evaluated 31 eyes of 31 patients aged ≤3.5 years, who had undergone secondary IOL implantation. The median absolute error (MedAE) was compared between axial length vergence formulas (Hoffer Q, Holladay I, SRK II, and SRK/T) and refractive vergence formulas (Lanchulev, Holladay R, Mackool, and Khan) as well as between formulas within the same vergence.
RESULTS: There was a significant difference (P = 0.010) between MedAE for axial length vergence formulas [1.19 Diopter(D)] and MedAE for refractive vergence formulas (2.48 D). The MedAE of axial length vergence formulas were comparable as to Hoffer (1.59 D), Holladay (1.27 D), SRK/T (1.23 D), and SRK II (1.30 D). Among refractive vergence formulas, Lanchulev (5.00 D) and Holladay R (2.51 D) had significantly larger MedAE as compared to Khan (2.06 D) and Mackool (2.15 D).
CONCLUSION: Axial length vergence formulas performed significantly better than refractive vergence formulas; however, axial length vergence formulas were comparable within the same vergence.