Comparison of AI-Automated and Manual Subfoveal Choroidal Thickness Measurements in an Elderly Population Using Optical Coherence Tomography.
СтатьяСравнительное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Zhou Wen-Da, Zhao Han-Qing, Geng Jia-Qi, Yang Yu-Hang, Dong Li, Zhang Rui-Heng, Wei Wen-Bin, Shao Lei
Аннотация
PURPOSE: To evaluate the agreement and correlation between manual and automated measurements of subfoveal choroidal thickness (SFCT) using enhanced depth imaging spectral-domain optical coherence tomography in an elderly population and to investigate the factors influencing measurement discrepancies.
METHODS: Based on the Beijing Eye Study, SFCT was measured manually using Heidelberg Eye Explorer software and automatically via a TransUNet-based deep learning model. Agreement between manual and automated SFCT measurements was assessed using Bland-Altman plots, intraclass correlation coefficients (ICC), and Pearson correlation coefficients.
RESULTS: Among 2896 participants, automated and manual measurements of SFCT demonstrated strong correlation (ICC = 0.971; 95% confidence interval [CI], 0.969-0.973; Pearson = 0.974, P < 0.001). Subgroup analyses showed similarly high correlation across participants aged ≥60 years (ICC = 0.954, Pearson = 0.974), aged <60 years (ICC = 0.971; Pearson = 0.953), with axial length ≥23 mm (ICC = 0.969; Pearson = 0.974), and axial length <23 mm (ICC = 0.959; Pearson = 0.963). Participants with SFCT <300 µm showed higher consistency (ICC = 0.942; Pearson = 0.944) compared to those with SFCT ≥300 µm (ICC = 0.867; Pearson = 0.868). Significant fixed and proportional biases were observed in all subgroups (P < 0.001), with manual measurements consistently lower than automated values.
CONCLUSIONS: Despite the presence of systematic biases, automated SFCT measurements showed excellent consistency and strong correlation with manual measurements across a large elderly population. These findings support the potential utility of AI-assisted SFCT measurement in clinical settings.
TRANSLATIONAL RELEVANCE: This study validates AI-based SFCT measurement in a large elderly cohort, enhancing diagnostic accuracy and bridging research with practice.
METHODS: Based on the Beijing Eye Study, SFCT was measured manually using Heidelberg Eye Explorer software and automatically via a TransUNet-based deep learning model. Agreement between manual and automated SFCT measurements was assessed using Bland-Altman plots, intraclass correlation coefficients (ICC), and Pearson correlation coefficients.
RESULTS: Among 2896 participants, automated and manual measurements of SFCT demonstrated strong correlation (ICC = 0.971; 95% confidence interval [CI], 0.969-0.973; Pearson = 0.974, P < 0.001). Subgroup analyses showed similarly high correlation across participants aged ≥60 years (ICC = 0.954, Pearson = 0.974), aged <60 years (ICC = 0.971; Pearson = 0.953), with axial length ≥23 mm (ICC = 0.969; Pearson = 0.974), and axial length <23 mm (ICC = 0.959; Pearson = 0.963). Participants with SFCT <300 µm showed higher consistency (ICC = 0.942; Pearson = 0.944) compared to those with SFCT ≥300 µm (ICC = 0.867; Pearson = 0.868). Significant fixed and proportional biases were observed in all subgroups (P < 0.001), with manual measurements consistently lower than automated values.
CONCLUSIONS: Despite the presence of systematic biases, automated SFCT measurements showed excellent consistency and strong correlation with manual measurements across a large elderly population. These findings support the potential utility of AI-assisted SFCT measurement in clinical settings.
TRANSLATIONAL RELEVANCE: This study validates AI-based SFCT measurement in a large elderly cohort, enhancing diagnostic accuracy and bridging research with practice.