OBJECTIVE

This paper presents the design of an electrically small circularly polarized (CP) 3 × 3 mm2 antenna system as an intraocular unit for retinal prosthesis application.

METHODS

The system is operating in ISM and ultra-wideband (UWB) bands to target high programmability of retina stimulation and recording, respectively. The electrical dimensions, including the ground plane, are λ0/41 × λ0/41 × λ0/191. Physical limitations of the antenna are discussed based on Hansen and Collin's limitations. The proposed wire patch antenna exhibits wideband characteristics by combining multiple modes of the patch antenna in the presence of an interface PCB circuit.

RESULTS

By loading polyimide encapsulated patch with stubs, dominant TM010 mode is combined with the higher order modes TM020-TM070 to exhibit wide -10 dB impedance bandwidth of 2-11 GHz. Annular rings and shorting pins in the ground plane provide CP radiation at 2.45, 5.8, and 8 GHz with 3-dB axial-ratio bandwidth of 0.3, 0.16, and 1.2 GHz, and far-field left hand circularly polarized (LHCP) gain of -18.4, -7.6, and -4.7 dBic, respectively, in broadside direction. A biocompatible antenna system is designed using Ansys HFSS in the presence of a detailed multilayer canonical eye model. Additionally, it is examined in an anatomical HFSS head model. Near and far-field electric field distribution is studied along with peak 1-g average specific absorption rate (SAR) calculations.

CONCLUSION

The proposed antenna is fabricated, and the performance, including coupled power from an external antenna, is measured in a custom made eye model including head phantom. A reasonable agreement is obtained between simulated and measured results.

SIGNIFICANCE: To generate an artificial vision, image perception capability could be improved with implantable UWB communication systems that feature particularly high data-rate and small size.

Для цитирования: Kaim Vikrant, Kanaujia Binod Kumar, Kumar Sachin, Choi Hyun Chul, Kim Kang Wook, Rambabu Karumudi. Electrically Small Circularly Polarized UWB Intraocular Antenna System for Retinal Prosthesis.. IEEE transactions on bio-medical engineering. 2022. DOI: 10.1109/tbme.2022.3171842.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.