Diabetes care · 2024

Efficacy and Safety of a Tubeless AID System Compared With Pump Therapy With CGM in the Treatment of Type 1 Diabetes in Adults With Suboptimal Glycemia: A Randomized, Parallel-Group Clinical Trial.

СтатьяРандомизированное исследованиеМногоцентровое исследованиеПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Renard Eric, Weinstock Ruth S, Aleppo Grazia, Bode Bruce W, Brown Sue A, Castorino Kristin, Hirsch Irl B, Kipnes Mark S, Laffel Lori M, Lal Rayhan A, Penfornis Alfred, Riveline Jean-Pierre, Shah Viral N, Thivolet Charles, Ly Trang T, OP5-003 Research Group

Аннотация

OBJECTIVE: To examine the efficacy and safety of the tubeless Omnipod 5 automated insulin delivery (AID) system compared with pump therapy with a continuous glucose monitor (CGM) in adults with type 1 diabetes with suboptimal glycemic outcomes.

RESEARCH DESIGN AND METHODS: In this 13-week multicenter, parallel-group, randomized controlled trial performed in the U.S. and France, adults aged 18-70 years with type 1 diabetes and HbA1c 7-11% (53-97 mmol/mol) were randomly assigned (2:1) to intervention (tubeless AID) or control (pump therapy with CGM) following a 2-week standard therapy period. The primary outcome was a treatment group comparison of time in range (TIR) (70-180 mg/dL) during the trial period.

RESULTS: A total of 194 participants were randomized, with 132 assigned to the intervention and 62 to the control. TIR during the trial was 4.2h/day higher in the intervention compared with the control group (mean difference 17.5% [95% CI 14.0%, 21.1%]; P < 0.0001). The intervention group had a greater reduction in HbA1c from baseline compared with the control group (mean ± SD -1.24 ± 0.75% [-13.6 ± 8.2 mmol/mol] vs. -0.68 ± 0.93% [-7.4 ± 10.2 mmol/mol], respectively; P < 0.0001), accompanied by a significantly lower time <70 mg/dL (1.18 ± 0.86% vs. 1.75 ± 1.68%; P = 0.005) and >180 mg/dL (37.6 ± 11.4% vs. 54.5 ± 15.4%; P < 0.0001). All primary and secondary outcomes were met. No instances of diabetes-related ketoacidosis or severe hypoglycemia occurred in the intervention group.

CONCLUSIONS: Use of the tubeless AID system led to improved glycemic outcomes compared with pump therapy with CGM among adults with type 1 diabetes, underscoring the clinical benefit of AID and bolstering recommendations to establish AID systems as preferred therapy for this population.