DOI 10.1007/s10384-022-00935-y

Медицинская библиотека · Japanese journal of ophthalmology · 2022

Agreement in the detection of chiasmal and postchiasmal visual field defects between imo binocular random single-eye test and Humphrey monocular test.

Sakamoto Mari, Sawamura Hiromasa, Aihara Makoto, Goseki Toshiaki, Ikeda Tetsuya, Ishikawa Hitoshi, Nakamura Makoto

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the ability of imo binocular random single-eye test (BRSET) to detect visual field (VF) defects due to chiasmal and postchiasmal lesions (C/PCLs) with a Humphrey Field Analyzer (HFA) monocular test.

STUDY

DESIGN

Prospective multicenter study

METHODS

This study enrolled 40 patients with C/PCLs and measured their VFs using both imo BRSET and HFA monocular test. The VFs were classified into three groups using the cluster criterion: 1) bitemporal group, 2) homonymous group, and 3) others. The agreement and correlation of VF results between imo and HFA were analyzed using the Bland-Altman plot and Spearman correlation coefficient.

RESULTS

The VFs of 34 patients were analyzed and classified. There were 13 patients in the bitemporal, 6 in the homonymous, and 15 in the others group. BRSET showed a significantly shorter test duration than HFA. The imo systematically yielded a lower sensitivity than HFA. The average sensitivity at each test location correlated well between the perimeters in all groups, with the correlation coefficients ranging from 0.89 to 0.98. Bland-Altman plots showed wider limits of agreement in the affected quadrants compared to the unaffected quadrants in the bitemporal and homonymous groups. The fixation loss rate did not differ between the perimeters, but there were significant differences in the false positive and false negative rates between perimeters.

CONCLUSION

BRSET detected VF defects due to C/PCLs as accurately as the HFA monocular test with a shorter test duration.

Для цитирования: Sakamoto Mari, Sawamura Hiromasa, Aihara Makoto, Goseki Toshiaki, Ikeda Tetsuya, Ishikawa Hitoshi, Nakamura Makoto. Agreement in the detection of chiasmal and postchiasmal visual field defects between imo binocular random single-eye test and Humphrey monocular test.. Japanese journal of ophthalmology. 2022. DOI: 10.1007/s10384-022-00935-y.