Agreement Between 10-2 and 24-2C Visual Field Test Protocols for Detecting Glaucomatous Central Visual Field Defects.

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Chakravarti Tutul, Moghadam Mohamad, Proudfoot James A, Weinreb Robert N, Bowd Christopher, Zangwill Linda M

Аннотация

PRECIS: Moderate to substantial agreement between 10-2 and 24-2C perimetry for detecting central field defects suggests that adding central test points to the 24-2 protocol may improve efficiency of visual field (VF) testing for glaucoma management.

PURPOSE: The purpose of this study was to assess agreement between Humphrey Visual Field Analyzer 10-2 and 24-2C test protocols for detecting glaucomatous defects in the central 10 degrees of the visual field (CVFDs).

MATERIALS AND METHODS: VFs from 165 eyes of 18 healthy individuals, 12 glaucoma suspects and 62 glaucoma patients who completed 10-2 and 24-2C VF testing protocols within 6 months were included. CVFDs on 10-2 and 24-2C (within the central 22 points) test grids required a cluster of 3 contiguous points with P<5%, 5%, and 1% or <5%, 2%, and 2% within a hemifield on the total deviation (TD) or pattern deviation (PD) plot. Cohen kappa (k) was used to assess agreement between 10-2 and 24-2C test grids in identifying CVFDs. Specificity of each testing strategy was assessed in VFs from healthy eyes.

RESULTS: CVFDs in suspect and glaucoma eyes were combined and reported as localized to superior, inferior or both hemifields based on TD and PD plots for 10-2 and 24-2C test grids. Moderate to substantial agreement was observed between 10-2 and 24-2C grids for detecting any CVFD from PD (k=0.551) and TD (k=0.651) plots. Specificity was high in healthy eyes ranging from 0.94 to 1.0 for both test protocols.

CONCLUSION: Substantial agreement for identifying CVFDs using the 24-2C and 10-2 protocols suggests that combining tests by adding central test points to the 24-2 test grid may supplant the need for 2 perimetry regimens for detecting central and peripheral glaucomatous VF damage.