The cardiac effects of hyperbaric oxygen at 243 kPa using inchamber echocardiography.

Gawthrope Ian C, Playford David A, King Benjamin, Brown Kathryn, Wilson Catherine, McKeown Barry

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The adverse effects of hyperbaric oxygen (HBO) on cardiac physiology are considered a potential hazard during the treatment of some patients. The haemodynamic effects of HBO are poorly understood and the incompatibility of electrical equipment inside the chamber has made assessment difficult. At Fremantle Hyperbaric Unit, we have modified an ultrasound machine (Logiq™ e) for safe use within the hyperbaric environment. The aim of this study was to evaluate the cardiac changes that occur during HBO using in-chamber transthoracic echocardiography (TTE) in subjects without evidence of active cardiac disease.

METHODS

Eleven patients and nine members of staff underwent comprehensive TTE examinations before and during HBO administered at a pressure of 243 kPa. The TTE examinations were reported by two independent cardiologists and statistically evaluated using paired Student's t-tests.

RESULTS

There was a significant decrease in heart rate during HBO (65 vs. 70 bpm on air at atmospheric pressure, P = 0.002) resulting in a decrease in cardiac output (5.3 vs. 5.9 L·min⁻¹, P = 0.003). Left ventricular outflow tract (LVOT) dimension was larger during HBO than baseline imaging (2.30 vs. 2.23 cm, P = 0.0003). LVOT velocity time integrals (VTI) decreased (19.9 vs. 21.5 cm, P = 0.009) and therefore a similar stroke volume was maintained (61 vs. 65 ml, P = 0.5). Ventricular and atrial volumes, intracardiac flows and minor valvular abnormalities were not significantly affected by HBO. No adverse cardiac effects were observed.

CONCLUSIONS

TTE can be safely performed within a hyperbaric chamber. Cardiac physiology is not adversely affected by HBO in individuals without active cardiac disease.

Для цитирования: Gawthrope Ian C, Playford David A, King Benjamin, Brown Kathryn, Wilson Catherine, McKeown Barry. The cardiac effects of hyperbaric oxygen at 243 kPa using inchamber echocardiography.. Diving and hyperbaric medicine. 2014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.