DOI 10.1097/01.mlg.0000172274.73365.11

Медицинская библиотека · The Laryngoscope · 2005

Shape-memory stapes prosthesis for otosclerosis surgery.

Knox Glenn W, Reitan Harlan

Клиническое исследованиеСравнительное исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to determine the efficacy of a shape-memory alloy, Nitinol, as a component of an improved stapes prosthesis.

STUDY

DESIGN

Prospective laboratory and clinical study to develop a Nitinol stapes prosthesis.

METHODS

Various diameters of Nitinol wire and temperature transition variants were analyzed with regard to ease of deformation, response to heating, and strength. The size and geometry of the closed hook was determined by measurement of 50 incus cadaver bones. Several heat sources for activating the shape memory were evaluated, including electrocautery, lasers, and warm water. Trial surgeries were then performed on human temporal bones in the laboratory. The closure characteristics of the Nitinol loop were studied. Magnetic resonance imaging (MRI) testing at 1.5 Tesla was performed to determine safety during MRI studies. Preliminary human subject trials were then instituted.

RESULTS

In all cases, a low heat condition was ample to activate the shape memory characteristics of the hook and return it to a closed position after it had been opened. Laser power was generally set well below the power needed for removing bone. The Nitinol loop closed snugly around the incus with application to the top of the hook with a low temperature laser setting. Almost any heat source was effective. MRI testing at 1.5 Tesla showed no movement of the prosthesis. Preliminary results in human subjects showed excellent air-bone closure. The Nitinol loop holds uniform contact around the incus.

CONCLUSIONS

The Nitinol piston greatly simplifies the stapedectomy procedure by taking the need for a hand operated instrument out of the surgeon's hands. Because of the nature of the Nitinol wire, it can never over-crimp. All these characteristics make the prosthesis advantageous for otosclerosis surgery.

Для цитирования: Knox Glenn W, Reitan Harlan. Shape-memory stapes prosthesis for otosclerosis surgery.. The Laryngoscope. 2005. DOI: 10.1097/01.mlg.0000172274.73365.11.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.