MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/00129492-200311000-00006

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2003

Retrospective analysis of early postoperative hearing results obtained after stapedotomy with implantation of a new titanium stapes prosthesis.

Zuur C L, de Bruijn A J G, Lindeboom R, Tange R A

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To evaluate the early postoperative hearing results of a new titanium stapes prosthesis (K-Piston) implanted in patients with otosclerosis.

STUDY

DESIGN

A retrospective analysis of preoperative and early postoperative hearing thresholds.

SETTING

One tertiary referral and teaching hospital.

PATIENTS: Eighteen men and 40 women, mean age 47 years, with otosclerosis.

INTERVENTION: Primary stapedotomy.

MAIN OUTCOME MEASURE: Main outcome measures were the mean gains in bone-conduction and air-conduction pure-tone thresholds, and pure-tone averages for different frequency combinations. Success and failure of the individual cases were presented using Amsterdam Hearing Evaluation Plots.

RESULTS

The overall postoperative air-bone gap for the frequency combination 0.5-1-2-4 kHz was 8.4 (standard deviation: 5.2) dB. In 79% of the patients the postoperative air-bone gap was less than 10 dB. Air-conduction improved even in higher frequencies, while the Carhart effect was not seen in most cases. In three patients a deterioration of bone-conduction was observed ranging from 11 to 16 dB sound pressure level (SPL), and in four patients the gain in air-conduction was insufficient (3-29 dB SPL) to close the preoperative air-bone gap to within 20 dB.

CONCLUSION

The new low-weight, full-titanium stapes prosthesis with its slight rough surface and its good mechanical stability and biocompatibility can safely and successfully restore the function of the middle ear when implanted in patients with otosclerosis.

Для цитирования: Zuur C L, de Bruijn A J G, Lindeboom R, Tange R A. Retrospective analysis of early postoperative hearing results obtained after stapedotomy with implantation of a new titanium stapes prosthesis.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2003. DOI: 10.1097/00129492-200311000-00006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.