Bicuspid Aortic Valve Anatomy and Relationship With Devices: The BAVARD Multicenter Registry.
СтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Tchetche Didier, de Biase Chiara, van Gils Lennart, Parma Radoslaw, Ochala Andrzej, Lefevre Thierry, Hovasse Thomas, De Backer Ole, Sondergaard Lars, Bleiziffer Sabine, Lange Rudiger, Kornowski Ran, Landes Uri, Norgaard Bjarne Linde, Biasco Luigi, Philippart Raphaël, Molina-Martin de Nicolas Javier, Mylotte Darren, Lemee Caroline, Dumonteil Nicolas, Van Mieghem Nicolas M
Аннотация
BACKGROUND: Sizing for transcatheter aortic valve implantation in bicuspid aortic valves (BAV) remains controversial.
METHODS AND RESULTS: The aim of the BAVARD (Bicuspid Aortic Valve Anatomy and Relationship With Devices) retrospective registry is to capture the sizing ratios used for transcatheter aortic valve implantation in BAV and analyze the second-generation prostheses geometry postimplantation. About 101 patients with BAV along with available pre- and post-transcatheter aortic valve implantation multidetector computed tomography were compared with 88 tricuspid aortic valves (TAV) patients. Preprocedural multidetector computed tomography diagnosed type 0 and type 1 BAV in, respectively, 12.9% and 86.1 % of BAV. At baseline, the ellipticity index was similar between BAV and TAV patients: 1.2±0.1 versus 1.2±0.1, P=0.09. The mean annular oversizing was, respectively, 1.14±0.04 and 1.04±0.04, P<0.001, in TAV and BAV patients. The mean prosthesis intercommissural distance, ratio was 1.03±0.1. The mean diameter of the prostheses at the annulus matched the mean perimeter-derived diameter of the aortic annulus at baseline with TAV (23.3±2.2 versus 23.6±1.9, P=0.4) and was smaller with BAV (24±2.8 versus 26.8±3.1, P<0.01), confirming 11% underexpansion in BAV. Finally, in situ, prosthesis diameter and ellipticity followed the same pattern, with stable values from the distal edge to 12 mm above, in both groups.
CONCLUSIONS: Second-generation prostheses similarly reshape the aortic annulus in TAV and BAV. Prostheses keep consistent diameters from distal edge to 12 mm in TAV and BAV. Prosthesis underexpansion is constantly observed in BAV. Annular-based sizing is accurate in BAV with minimal oversizing. The intercommissural distance, 4 mm above the annulus, could be integrated in gray zones.
CLINICAL TRIAL REGISTRATION: URL: https://www.clinicaltrials.gov . Unique identifier: NCT03495050.
METHODS AND RESULTS: The aim of the BAVARD (Bicuspid Aortic Valve Anatomy and Relationship With Devices) retrospective registry is to capture the sizing ratios used for transcatheter aortic valve implantation in BAV and analyze the second-generation prostheses geometry postimplantation. About 101 patients with BAV along with available pre- and post-transcatheter aortic valve implantation multidetector computed tomography were compared with 88 tricuspid aortic valves (TAV) patients. Preprocedural multidetector computed tomography diagnosed type 0 and type 1 BAV in, respectively, 12.9% and 86.1 % of BAV. At baseline, the ellipticity index was similar between BAV and TAV patients: 1.2±0.1 versus 1.2±0.1, P=0.09. The mean annular oversizing was, respectively, 1.14±0.04 and 1.04±0.04, P<0.001, in TAV and BAV patients. The mean prosthesis intercommissural distance, ratio was 1.03±0.1. The mean diameter of the prostheses at the annulus matched the mean perimeter-derived diameter of the aortic annulus at baseline with TAV (23.3±2.2 versus 23.6±1.9, P=0.4) and was smaller with BAV (24±2.8 versus 26.8±3.1, P<0.01), confirming 11% underexpansion in BAV. Finally, in situ, prosthesis diameter and ellipticity followed the same pattern, with stable values from the distal edge to 12 mm above, in both groups.
CONCLUSIONS: Second-generation prostheses similarly reshape the aortic annulus in TAV and BAV. Prostheses keep consistent diameters from distal edge to 12 mm in TAV and BAV. Prosthesis underexpansion is constantly observed in BAV. Annular-based sizing is accurate in BAV with minimal oversizing. The intercommissural distance, 4 mm above the annulus, could be integrated in gray zones.
CLINICAL TRIAL REGISTRATION: URL: https://www.clinicaltrials.gov . Unique identifier: NCT03495050.
Предметные рубрики MeSH
AgedAged, 80 and overAortic ValveAortic Valve StenosisBicuspid Aortic Valve DiseaseEuropeFemaleHeart Valve DiseasesHeart Valve ProsthesisHumansIsraelMaleMultidetector Computed TomographyProsthesis DesignRegistriesRetrospective StudiesRisk AssessmentRisk FactorsTime FactorsTranscatheter Aortic Valve ReplacementTreatment Outcome